Table of Contents

Advertisement

Quick Links

Planmeca Sovereign
installation manual

Advertisement

Table of Contents
loading

Summary of Contents for Planmeca Sovereign

  • Page 1 Planmeca Sovereign ™ installation manual...
  • Page 3: Table Of Contents

    Patient area ........................8 Fire stopping of the installation site ................9 UNIT CONFIGURATION ................10 Planmeca Sovereign with OP delivery arm and hanging tube instrument holders 11 Planmeca Sovereign with side delivery arm ..............12 OPERATIONS PRIOR TO THE CUSPIDOR ATTACHMENT ......14 Using the support for floor tubing .................
  • Page 4 IEC364 - equipment is used according to the operating instructions Planmeca pursues a policy of continual product development. Although every effort is made to produce up-to-date product documentation this publication should not be regarded as an infallible guide to current specifications. We re- serve the right to make changes without prior notice.
  • Page 5: Introduction

    The information given in this manual should be taken as a general guide for proper in- stallation of the Planmeca Sovereign dental unit. However, in the cases that this infor- mation is contradictory to any local or national building regulations, the building regulations must always override this manual.
  • Page 6: Contents Of The Packing

    CONTENTS OF THE PACKING CONTENTS OF THE PACKING Cuspidor covers, cuspidor bottom plate, legrest, seat adapter bottom plate, tray table. Monitor arm Delivery arm Cuspidor Assistant element Graphical user interface Foot control Installation manual 2 Planmeca Sovereign...
  • Page 7: Associated Documentation

    Verify the operation of the installed unit according to the User’s manual. You will also need the Planmeca Sovereign wiring diagram (publication number D0001057) during the installation. All the documentation listed above is supplied with the unit. The operating light has also its own installation manual.
  • Page 8: Pre-Installation Requirements

    CAUTION Do not connect a Multiple Portable Socket-Outlet or extension cord to the sys- tem. Arm movements Dental unit with 3- or 5-function pylon REFERENCE JOINT for balanced arm. Inner and outter joint of balanced arm. Installation manual 4 Planmeca Sovereign...
  • Page 9: Basement

    Accepted limits for unit transportation are the following: Temperatures: -20 °C to +60 °C (-4 °F to 140 °F) Relative humidity: 5% RH to 95% RH “Absence of condensation” Air pressure: 700 hPa to 1060 hPa (10 psi to 15 psi) Installation manual Planmeca Sovereign 5...
  • Page 10: Plumbing Requirements

    Avoid using sharp bends or joints in the suction line. Seal all joints to avoid air leakage. Care should be taken to avoid condensation, particularly when the pipe is exposed to different ambient temperatures. Moisture traps must be used if necessary. Installation manual 6 Planmeca Sovereign...
  • Page 11: Electrical Requirements

    The idle (instruments in holders) power consumption is 100 W. Maximum power consumption is 1000 W. Residual-current device The dental unit must be equipped with a residual-current device (RCD), which is also known as ground fault interrupter (GFI). Installation manual Planmeca Sovereign 7...
  • Page 12: Patient Area

    NOTE The monitor installed to the system must fulfil IEC 60950 requirements. NOTE Keyboard and mouse connected to external PC as well as Planmeca ProSensor’s power supply and Planmeca ProX generator assembly must be located outside the patient ar- NOTE Connect only Planmeca specified devices.
  • Page 13: Fire Stopping Of The Installation Site

    Fill the floor opening with fire stop sealing compound (e.g. FIREPRO M799 GPG Fire Stop sealing compound). Add waterproof material (e.g. Loctite 5366) above the compound and into the cable pipe. The figure below is an example (installation to a ready-made concrete floor). Installation manual Planmeca Sovereign 9...
  • Page 14: Unit Configuration

    UNIT CONFIGURATION UNIT CONFIGURATION Planmeca Sovereign with OP delivery arm and balanced instrument arms OP delivery arm with balanced instrument arms Monitor Planmeca Rotating bowl ProX assembly Graphical Assistant User element SingLED Interface operating (GUI) light Tray table Patient chair...
  • Page 15: Planmeca Sovereign With Op Delivery Arm And Hanging Tube Instrument

    OP delivery arm and hanging tube instrument holders Monitor SingLED operating light OP delivery arm with hanging tube (HT) instrument holders Graphical User Interface Tray table (GUI) Assistant element Bowl Patient chair Cuspidor Foot control Installation manual Planmeca Sovereign 11...
  • Page 16: Planmeca Sovereign With Side Delivery Arm

    UNIT CONFIGURATION Planmeca Sovereign with side delivery arm Monitor Patient chair Bowl Tray table Graphical Assistant Foot control User Cuspidor element Interface (GUI) Side delivery arm with hanging tube (HT) instrument holders Installation manual 12 Planmeca Sovereign...
  • Page 17 UNIT CONFIGURATION The following picture presents the Planmeca Sovereign with side delivery arm and equipped with the optional surgical and ortho tray table. OP delivery arm Surgical and ortho tray table Assistant element Bowl Patient chair Monitor Graphical User Interface...
  • Page 18: Operations Prior To The Cuspidor Attachment

    Refer to section 5.2 “Attaching the installation plate to the floor” on page 18. The dental unit will be positioned according to the instructions given in installation pattern supplied with the dental unit. Installation manual 14 Planmeca Sovereign...
  • Page 19 VGA cable/ power supply (optional) Intra cable Assistant call/ Door opening cable 4x0.75mm2 Video cable Oral X-ray cable Connection area for electricity and telecommunication SUCTION LINE AIR SUPPLY WATER SUPPLY DRAIN LINE Installation manual Planmeca Sovereign 15...
  • Page 20: Using The Support For Floor Tubing

    (2). Place the plywood template assembly over the tube support (3). Make sure that the wood blocks that shows the security area do not hit the tubes/cables (4). Attach the plywood assembly to the support plate screws with attachment nuts (5). 60mm Installation manual 16 Planmeca Sovereign...
  • Page 21 (3). Attach the tubes/cables to the support plate with e.g, wire (4). Cast the 80 mm (3.1 in.) thick floor casting. Cut the tubes that come from the floor so that they extend 25…50 mm (1…2 in.) from the floor (1). Installation manual Planmeca Sovereign 17...
  • Page 22: Attaching The Installation Plate To The Floor

    Attach the plate to the floor with 6x40 torx screws (3). Attach the base attachment screws to the installation plate (4). Install the cuspidor to the plate according to the instructions given in section 6 “INSTALLING THE CUSPIDOR”. Installation manual 18 Planmeca Sovereign...
  • Page 23 Attach the plate to the floor with 6x40 torx screws (3). Attach the base attachment screws to the installation plate (4). Install the cuspidor to the plate according to the instructions given in section 6 “INSTALLING THE CUSPIDOR”. Installation manual Planmeca Sovereign 19...
  • Page 24: Installing The Cuspidor

    The base cover attachment mechanism is located inside the cuspidor. Rotate the mechanism screw using a 6 mm Allen key to move the mechanism arm and to release the cover attachment (1). Pull the cover away from its position (2). Installation manual 20 Planmeca Sovereign...
  • Page 25 (right-handed/left-handed)”. Drill 14 mm (0.55 in.) holes until the required depth of 55 mm (2.2 in.). Clean the dust caused by the drilling from the holes with a vacuum cleaner or with pressure air. Installation manual Planmeca Sovereign 21...
  • Page 26 Move the cuspidor to its place. Position the dental unit according to the instructions given in installation pattern supplied with the dental unit. Note, that the openings on the base are in the direction of the chair legrest. Installation manual 22 Planmeca Sovereign...
  • Page 27 INSTALLING THE CUSPIDOR Insert the lock washers to the anchors (1) and rotate them to the position shown on the figure below (2). Insert the washers (3) and nuts (4). Installation manual Planmeca Sovereign 23...
  • Page 28: Dental Units With Fixed Base: Changing The Cuspidor Installation Position (Right-Handed/Left-Handed)

    Change the cuspidor position before removing the transport handles. CAUTION Aways rotate the cuspidor from right-handed to left-handed position COUN- TERCLOCKWISE and vice versa. DO NOT ROTATE THE CUSPIDOR IN AREA MARKED TO THE FIGURE ABOVE. Rotate the cuspidor up 240°. Installation manual 24 Planmeca Sovereign...
  • Page 29 INSTALLING THE CUSPIDOR Cuspidor In the positions CR2 and CL2 the cuspidor is attached to the base with five metal hooks and ten attachment screws. Installation manual Planmeca Sovereign 25...
  • Page 30 INSTALLING THE CUSPIDOR In the figure below the base is shown from below. The arrows indicate the positions of the metal hooks. Installation manual 26 Planmeca Sovereign...
  • Page 31 INSTALLING THE CUSPIDOR In positions CR1 and CL1 six screws attach the cuspidor to the base (four on the frame and two beside the cuspidor). Installation manual Planmeca Sovereign 27...
  • Page 32 Unscrew the screws using a 4 mm Allen key (1) and rotate the cuspidor (2). Position CR1 In CR2 position attach the cuspidor to the base with metal hooks and attachment screws (1). Position CR2 Installation manual 28 Planmeca Sovereign...
  • Page 33 In CR2 position attach the cuspidor to the base with metal hooks and attachment screws (1). Position CL2 In position CL1 attach the cuspidor to the base with six attachment screws (four on the frame and two beside the cuspidor). Position CL1 Installation manual Planmeca Sovereign 29...
  • Page 34 Unscrew the seat rotation locking screw (1). Turn the seat to the desired position (2). Attach the rotation locking screw (1). After changing the cuspidor installation position you also have to change the patient chair position 180° (±15°) (3). 15° 15° Installation manual 30 Planmeca Sovereign...
  • Page 35: Removing The Transport Handle And Wheels

    Loosen the transport handle attachment screws and remove the handle. Do not loosen the screws too much, otherwise the attachment nuts will drop into the cuspidor. Loosen the attachment screws of the wheels (1) and detach them from below the cuspidor (2). Installation manual Planmeca Sovereign 31...
  • Page 36: Checking The Vertical Position Of The Cuspidor

    The vertical position of the cuspidor is adjusted with adjustment plates supplied with the unit (2). Loosen the adjustment nuts and place the plate(s) between the cuspidor base and floor until the cuspidor is vertical. Re-check the cuspidor position and add more plate(s), if needed. Installation manual 32 Planmeca Sovereign...
  • Page 37 The vertical position of the cuspidor is adjusted with adjustment plates supplied with the unit (3). Loosen the adjustment nuts and place the plate(s) between the cuspidor base and floor until the cuspidor is vertical. Re-check the cuspidor position and add more plate(s), if needed. Installation manual Planmeca Sovereign 33...
  • Page 38: Connecting The Service Tubes

    WATER SUPPLY PIPE Water pressure 300-900kPa. Air pressure 550-900kPa. Copper pipe Ø12/10mm (0.5"/0.4") Copper pipe Ø12/10mm (0.5"/0.4") At the pipe end a nipple At the pipe end a nipple with female thread R1/4" with female thread R1/4" Installation manual 34 Planmeca Sovereign...
  • Page 39 Secure the tubes with attachment rings. SUCTION DRAIN SUCTION LINE PIPE DRAIN LINE PIPE Plastic pipe Ø50/46mm (2.0"/1.8") Plastic pipe Ø50/46mm (2.0"/1.8") (or copper Ø35/32mm (1.4"/1.3")) (or copper Ø35/32mm (1.4"/1.3")) Re-attach the suction and drain tubes coming from the cuspidor. Installation manual Planmeca Sovereign 35...
  • Page 40 INSTALLING THE CUSPIDOR Secure the water and air tubes to the base with a cable tie and cable tie holder. Installation manual 36 Planmeca Sovereign...
  • Page 41: Connecting The Cables

    VGA cable/ power supply (optional) Intra cable Assistant call/ Door opening cable 4x0.75mm2 Video cable Oral X-ray cable Connection area for electricity and telecommunication SUCTION LINE AIR SUPPLY WATER SUPPLY DRAIN LINE Installation manual Planmeca Sovereign 37...
  • Page 42 INSTALLING THE CUSPIDOR Unscrew two mains terminal cover attachment screws using 2 mm Allen key and remove the cover. Installation manual 38 Planmeca Sovereign...
  • Page 43 INSTALLING THE CUSPIDOR Unscrew the two attachment screws of the mains terminal shield (1) and turn the shield away from above the mains terminal (2). Installation manual Planmeca Sovereign 39...
  • Page 44 Notice that the mains voltage is always present at the mains terminal even, if the unit is switched off from its own mains switch. If required, connect the suction motor, door open and assistant call cables to the respective terminals. Shorten the cable if necessary. Installation manual 40 Planmeca Sovereign...
  • Page 45 XCAN RS-232 Isolation Monitor-DC Ethernet Isolation VGA Ext. Cable S-Video Intra NOTE See also figure on section 19 “CONNECTIONS” on page 181. Place the mains terminal cover to its position and tighten two attachment screws. Installation manual Planmeca Sovereign 41...
  • Page 46: Installing The Separate Connection Case (Optional)

    Mark the positions of the case attachment screws to the floor. If the floor is made of wood or similar, use the installation plate (refer to section “Installation plate with separate connection case”. Installation manual 42 Planmeca Sovereign...
  • Page 47 INSTALLING THE SEPARATE CONNECTION CASE (OPTIONAL) Drill attachment holes and attach the case attachment plates to the floor. Measure the distance between the base outer edge and case cover attachment screw opening located on the plate. Installation manual Planmeca Sovereign 43...
  • Page 48 INSTALLING THE SEPARATE CONNECTION CASE (OPTIONAL) Mark the attachment screw position to the case cover. Attach the case cover to its position with the self drilling screw. Detach the cover. Installation manual 44 Planmeca Sovereign...
  • Page 49 5 “OPERATIONS PRIOR TO THE CUSPIDOR ATTACHMENT” on page 14. Route the cables and tubes into the unit. Secure the cables and tubes with attachment strip. Route the water and air tubes from below the attachment plate. Installation manual Planmeca Sovereign 45...
  • Page 50 Connect the tubes according to the instructions given in section 6.5 “Connecting the service tubes” on page 34. Connect the cables according to the instructions given in section 6.6 “Connecting the cables” on page 37. Installation manual 46 Planmeca Sovereign...
  • Page 51 INSTALLING THE SEPARATE CONNECTION CASE (OPTIONAL) Connect all the cables and attach the mains terminal cover and the plastic shield. Attach the suction and drain tubes coming from the cuspidor and secure them with attachment strip. Installation manual Planmeca Sovereign 47...
  • Page 52: Installing The Pylon And Arms

    Refer to the section 6.4 “Checking the vertical position of the cuspidor” on page 32. The pylon adjustment cannot be corrected with the pylon attachment plate adjustment. Check the sideways position of the pylon attachment plate using a spirit level. Installation manual 48 Planmeca Sovereign...
  • Page 53 INSTALLING THE PYLON AND ARMS Check the position of the pylon attachment plate in depth with a spirit level. Installation manual Planmeca Sovereign 49...
  • Page 54 The pylon position in depth is adjusted with a screw located behind the attachment plate (use 5 mm Allen key). You can reach the screw from behind the lower edge of the plate (3). After adjustment tighten the locking screw. Installation manual 50 Planmeca Sovereign...
  • Page 55 Make sure that the attachment screws are loosened so that the they go into the openings on the pylon. Do not loosen the screws too much, otherwise the attachment nuts will drop into the cuspidor. Lift the pylon to its position as shown on the figure below. Installation manual Planmeca Sovereign 51...
  • Page 56 Tighten the four pylon attachment screws using a 10 mm Allen key. Check the sideways position of the pylon as well as the position of the pylon in depth using a spirit level. Adjust the pylon attachment plate position, if needed. Installation manual 52 Planmeca Sovereign...
  • Page 57: Installing The Rotating Op Delivery Arm

    Installing the rotating OP delivery arm Attach the lower pylon according to the instructions given in section 8.1 “Installing the pylon” on page 48. Slide the joint support plate to the pylon. Push the OP delivery arm to its position. Installation manual Planmeca Sovereign 53...
  • Page 58 After attaching the screws tighten them all. Remove the OP delivery arm transport plate and the OP delivery arm middle joint transport plate as described in section 8.3 “Installing the fixed OP delivery arm” on page 65. Installation manual 54 Planmeca Sovereign...
  • Page 59 INSTALLING THE PYLON AND ARMS Slide the joint support plate to the upper pylon. Place the upper pylon to the OP delivery arm joint so that the stopper arch (black arrow) is towards patient chair. Installation manual Planmeca Sovereign 55...
  • Page 60 The pylon is attached to the OP delivery arm with eight attachment screws and washers (four on both sides). Attach the screws to the both sides using a 6 mm Allen key. Tighten all screws. Installation manual 56 Planmeca Sovereign...
  • Page 61 INSTALLING THE PYLON AND ARMS Attach the upper pylon cover with four attachment screws using a 10 Torx key, Install the Planmeca SingLED operating light at this stage. Refer to the section 9.6 “Installing the Planmeca SingLED operating light” on page 110.
  • Page 62 Install the monitor arm to the pylon at this stage. Refer to the section 9.3 “Installing the monitor to the pylon” on page 93. NOTE: route the monitor cables through the OP delivery arm joint. Attach the lower pylon cover with four attachment screws using a 10 Torx key Installation manual 58 Planmeca Sovereign...
  • Page 63 Attach the arm joint side covers to their positions. The covers are attached with three screws: to the joint with two screws (white arrows) and to the arm with one screw. Use a 3 mm Allen key. Installation manual Planmeca Sovereign 59...
  • Page 64 Bend the support lugs inside with e.g. Allen key and press the support to its position. Connect the arm cables and tubes according to the instructions given in section “Connecting the OP delivery arm cables and tubes” on page 70. Installation manual 60 Planmeca Sovereign...
  • Page 65 INSTALLING THE PYLON AND ARMS Push the back cover attachment screw through the cover opening into the groove as shown on the figure below. Attach the back cover to its position. Installation manual Planmeca Sovereign 61...
  • Page 66 If needed, remove the upper cover and adjust the lower cover position so that the cover fit to each other. Adjust the vertical position as follows. Loosen four screws which attach the angle plate on the arm side using a 3 mm Allen key and move the cover. Tighten the screws. Installation manual 62 Planmeca Sovereign...
  • Page 67 Check that there is approx. 3 mm gap between the lower cover and middle joint upper cover. Attach the other OP delivery arm covers according to the instructions given in section 8.6 “Attaching the OP delivery arm covers” on page 81. Installation manual Planmeca Sovereign 63...
  • Page 68 INSTALLING THE PYLON AND ARMS Attach the pylon covers according to the instructions given in section 10 “ATTACHING THE PYLON COVERS” on page 122. Installation manual 64 Planmeca Sovereign...
  • Page 69: Installing The Fixed Op Delivery Arm

    Installing the fixed OP delivery arm Attaching the OP delivery arm The OP delivery arm is attached to the pylon openings marked with lower arrows (1). The monitor arm is attached to openings marked with upper arrows (2). Installation manual Planmeca Sovereign 65...
  • Page 70 Lift the OP delivery arm to its position. The arm attachment screws must go into the openings on the pylon as shown on the figure below. Make sure that the attachment screws are loosened so that the they go into the openings on the pylon. Lower attachment screws Installation manual 66 Planmeca Sovereign...
  • Page 71 (1). Tighten the attachment screws using a 6 mm Allen key (2). Make sure that the arm cables are secured with cable ties. Attach the OP delivery arm grounding cable to the pylon using a 3 mm Allen key (1). Installation manual Planmeca Sovereign 67...
  • Page 72 Bend the support lugs inside with e.g. Allen key and press the support to its position. Unscrew the OP delivery arm transport plate attachment screws (first the lower two, then the upper two) and remove the transport plate. Installation manual 68 Planmeca Sovereign...
  • Page 73 INSTALLING THE PYLON AND ARMS Loosen the OP delivery arm middle joint transport plate attachment screws (1) and remove the transport plate (2). Installation manual Planmeca Sovereign 69...
  • Page 74 INSTALLING THE PYLON AND ARMS Connecting the OP delivery arm cables and tubes Unscrew the attachment screw using a 2.5 mm Allen key (1) and remove the electronics control box cover (2). Installation manual 70 Planmeca Sovereign...
  • Page 75 INSTALLING THE PYLON AND ARMS Route the OP delivery arm cables to the electronics control box and secure them with cable clamps as shown on the figure below. Installation manual Planmeca Sovereign 71...
  • Page 76 INSTALLING THE PYLON AND ARMS Connect the cables to connectors on the UPSU PCB as shown on the figure below (refer also to the Sovereign wiring diagram). Route the arm tubes into the cuspidor and behind the suction system attachment plate to the block.
  • Page 77 OP-Arm SELV Cable Assistant-Arm SELV Cable If instrument console is equipped with Intraoral video camera, connect the S-video cable according to the instructions given in section 9.3 “Installing the monitor to the pylon” on page Installation manual Planmeca Sovereign 73...
  • Page 78: Installing The Balanced Arm Instrument Console

    Installing the balanced arm instrument console Attach the instrument console to the console bracket using the support plate. Tighten the attachment screws using a 3 mm Allen key. Route the arm cables and tubes to the instrument console. Installation manual 74 Planmeca Sovereign...
  • Page 79 INSTALLING THE PYLON AND ARMS Connect the arm water (1), air (2) and return water (3) tubes as shown on the figure below. Installation manual Planmeca Sovereign 75...
  • Page 80 Return water Air IN S-Video LOCAN SELV EMEIO EMEIO Micro-Fit 4 Micro-Fit 4 SELV LOCAN Mini-Fit 6 Mini-Fit 4 P4 P3 Cover the water tube connections with protective sleeves as shown on the figure above. Installation manual 76 Planmeca Sovereign...
  • Page 81 Secure the cables with cable ties (1). Secure the Selv cable connected to connector P7 with one attachment screw (2). Make sure that the Selv cable leans on the screw located on the ICON PCB (3). Installation manual Planmeca Sovereign 77...
  • Page 82: Installing The Ht Instrument Console

    Cut the cable tie that binds the arm cables and tubes. Route the cables and tubes through the arm opening as shown on the figure below. Lift the instrument console to its position so that the pin on the console goes into the arm groove (arrow on the figure below). Installation manual 78 Planmeca Sovereign...
  • Page 83 Attach the instrument console to the console arm with four attachment screws using a 3 mm Allen key. Route the arm cables and tubes to the instrument console. Connect the arm water (1), air (2) and return water (3) tubes as shown on the figure below. Installation manual Planmeca Sovereign 79...
  • Page 84 Connect the cables as shown on the figure below. Cover the water tube connections with protective sleeves. Secure the cables with cable ties (1). Secure the Selv cable connected to connector P7 with one attachment screw (2). Installation manual 80 Planmeca Sovereign...
  • Page 85: Attaching The Op Delivery Arm Covers

    Attaching the OP delivery arm covers Fixed OP delivery arm: Attach the console bracket cover to its position with four attachment screws (1). Attach the OP-delivery arm middle joint cover (2) with one attachment screw (1). Installation manual Planmeca Sovereign 81...
  • Page 86 INSTALLING THE PYLON AND ARMS Press the OP delivery arm to its lowest position and slide the cover to below the joint cover (1) and press the cover to its position (2). Installation manual 82 Planmeca Sovereign...
  • Page 87 Attach the console bracket lower cover as follows. Place the cover to its position so that the lug on the arm goes into the cover groove. Attach the arm with three screws using a 3 mm Allen key. Attach the instrument console upper cover using a 3 mm Allen key. Installation manual Planmeca Sovereign 83...
  • Page 88 INSTALLING THE PYLON AND ARMS HT instrument console covers Attach the console joint cover with two attachment screws using a 3mm Allen key. Attach the instrument console upper cover using a 3 mm Allen key Installation manual 84 Planmeca Sovereign...
  • Page 89: Attaching/Detaching The Graphic User Interface (Gui)

    Push the GUI arm into the attachment opening in position where the GUI arm is rotated approx. 30° forwards as shown on the figure below. Attach the GUI to the console by rotating the GUI arm clockwise until you hear a click sound. 30° & Installation manual Planmeca Sovereign 85...
  • Page 90 Press the locking button located underside the GUI arm (1) and turn the GUI arm approx. approx. 30° clockwise (right side) (2) or counterclockwise (left side). Detach the GUI by pulling the arm out from the instrument console (3). & Installation manual 86 Planmeca Sovereign...
  • Page 91: Installing The Pylon And Arm Accessories

    INSTALLING THE PYLON AND ARM ACCESSORIES INSTALLING THE PYLON AND ARM ACCESSORIES Attaching tray to the balanced arm instrument console The mounting arm is attached to the instrument console by pushing it to its position. Installation manual Planmeca Sovereign 87...
  • Page 92 The friction of the tray arm movements can be adjusted. Loosen the joint attachment screw using a 3 mm Allen key (1). Rotate the plate manually and move the tray arm (2). Hold the plate in position with pliers (3) and tighten the attachment screw (4). Installation manual 88 Planmeca Sovereign...
  • Page 93 INSTALLING THE PYLON AND ARM ACCESSORIES The angle of the tray table can be adjusted as follows. Loosen the securing screws (1) and adjust the angle (3) with four adjustment screws (2). Tighten the securing screws. Installation manual Planmeca Sovereign 89...
  • Page 94: Attaching Rotating Tray To The Ht Instrument Console

    Attach the tray mounting arm to the instrument console by rotating the tray mounting arm counterclockwise (when the tray table is placed on the left side of the console) or clockwise (when the tray table is placed on the right side of the console) (2). Installation manual 90 Planmeca Sovereign...
  • Page 95 (when the tray table is placed on the right side of the console). Detach the tray by pulling the arm out from the instrument console (3). Installation manual Planmeca Sovereign 91...
  • Page 96 Loosen the tray joint locking screw using a 1.5 mm Allen key (1). Adjust the rotational friction of tray table by turning the attachment plate with pliers (2). Turning the plate clockwise increases the fric- tion. Tighten the locking screw. Installation manual 92 Planmeca Sovereign...
  • Page 97: Installing The Monitor To The Pylon

    INSTALLING THE PYLON AND ARM ACCESSORIES Installing the monitor to the pylon NOTE This section applies to both Planmeca white (HFN17F) and Planmeca Vision monitors. Any installation methods differentiating from each other are specified in text. NOTE The PC connected to the monitor must be:...
  • Page 98 Secure the cables with cable ties (1). Attach the grounding cable to the pylon (2). The monitor cables are attached to the monitor arm at the factory with a cable tie. Cut the cable tie. Installation manual 94 Planmeca Sovereign...
  • Page 99 (3). Attach the lower attachment bushings to the monitor as shown on the figure below (4,5 and 6). Tighten the attachment screws of the upper bushings. Planmeca Vision monitor Planmeca white monitor Installation manual Planmeca Sovereign 95...
  • Page 100 INSTALLING THE PYLON AND ARM ACCESSORIES Planmeca Vision monitor: attaching the covers and connecting the cables Cut an opening to the monitor back cover as shown on the figure below. Attach the monitor back cover to its position as follows. Loosen the lower attachment screws (1) and turn the monitor slightly upwards (2).
  • Page 101 INSTALLING THE PYLON AND ARM ACCESSORIES Attach the lower cover support piece to the attachment plate. Place the upper cover to its position and secure tentatively it with two attachment screws using a 3 mm Allen key. Installation manual Planmeca Sovereign 97...
  • Page 102 3 mm Allen key. Fit the covers to their positions and tighten the four attachment screws Route the cables into the cuspidor. Connect the power cable to the cable coming from the UPSU PCB. Installation manual 98 Planmeca Sovereign...
  • Page 103 S-video (6) and DisplayPort (7) extension cables to the computer. DisplayPort, Male Monitor arm S-Video Dentists arm Double isolation Double isolation DisplayPort, DisplayPort, Male Female DisplayPort, S-Video Male S-Video NOTE See also figure on section 19 “CONNECTIONS” on page 181. Installation manual Planmeca Sovereign 99...
  • Page 104 INSTALLING THE PYLON AND ARM ACCESSORIES Planmeca white monitor: attaching the covers and connecting the cables Place the upper cover to its position (1) and secure tentatively it with two attachment screws (2) using a 3 mm Allen key. Place the lower attachment cover (3) to its position and secure it with two attachment screws using a 3 mm Allen key (4).
  • Page 105 (4, 5, 6). Route the cables out from the base and attach the S-video (7) and VGA (8) extension cables to the computer. Monitor arm S-Video Dentists arm MiniDin MiniDin S-Video S-Video NOTE See also figure on section 19 “CONNECTIONS” on page 181. Installation manual Planmeca Sovereign 101...
  • Page 106 If needed, you can adjust the friction of the monitor angle movement as shown on the figures below. Remove the upper cover of the monitor arm. Adjust the friction with a 2.5 mm Allen key. Adjust the two screws equally (1). Tightening the screw increases the friction (2) Installation manual 102 Planmeca Sovereign...
  • Page 107: Installing The Planmeca White Monitor To The Ht Console

    INSTALLING THE PYLON AND ARM ACCESSORIES Installing the Planmeca white monitor to the HT console NOTE The PC connected to the monitor must be: - IEC-approved (CE marked) - located outside the patient area and - protectively earthed. CAUTION Do not connect items which are not specified as part of the system.
  • Page 108 INSTALLING THE PYLON AND ARM ACCESSORIES Attach the power cable to the VGA cable with tape as shown on the figure below. Route the cables through the monitor arm. Installation manual 104 Planmeca Sovereign...
  • Page 109 Attach the lower attachment bushings as shown on the figure below. Tighten the attach- ment screws of the upper bushings. Installation manual Planmeca Sovereign 105...
  • Page 110 Connect the power cable to the monitor connector. Pull the excess monitor cables into the console. Carefully place the cables into the arm adapter groove. Attach the adapter bottom plate with three attachment screws. Installation manual 106 Planmeca Sovereign...
  • Page 111 PCB. Route the VGA cable into the base and connect it to the VGA extension cable. Isolate the con- nection and place the excess VGA arm cable into the insulator. Attach the VGA extension cable to the computer. Attach the console cover. Installation manual Planmeca Sovereign 107...
  • Page 112: Installing The Sterile Water System To The Instrument Console

    The Sterile water system cannot be installed to the OP delivery arm with HT console if also the LCD monitor is installed. Remove the console cover. Attach the pump to the console with 2 attachment screws using a 3 mm Allen key. Installation manual 108 Planmeca Sovereign...
  • Page 113 Make sure that the pump motor cable is connected to the ICON PCB connector P14 (S-pump motor). Connect the cable to the pump. Attach the water bag holder to below the instrument console with one attachment screw using a 3 mm Allen key. Attach the console cover. Installation manual Planmeca Sovereign 109...
  • Page 114: Installing The Planmeca Singled Operating Light

    INSTALLING THE PYLON AND ARM ACCESSORIES Installing the Planmeca SingLED operating light NOTE Install the monitor arm before installing the operating light. Route the light cable through the pylon top bracket to outer side of the pylon. Insert the three washers to the operating light arm shaft and push the arm to the bracket opening.
  • Page 115: Installing Zeiss Opmi Pico Microscope

    The microscope arm is attached to the pylon openings marked with arrows (1). Lift the microscope arm to its position. Make sure that the attachment screws are loosened so that the they go into the openings on the pylon. Installation manual Planmeca Sovereign 111...
  • Page 116 Attach the microscope arm to its position. First slide the cable support plate under the lower attachment screws. Tighten the attachment screws. Secure the cables with cable ties. Assemble the microscope according to the manufacturer’s instructions. Installation manual 112 Planmeca Sovereign...
  • Page 117 Route the cables through the stand arm opening and place the stand arm on the microscope arm shaft. Secure the stand arm in its position with washer and two c-clips. Be careful not to overstretch the c-clips when mounting them. Installation manual Planmeca Sovereign 113...
  • Page 118 Connect the S-video cables (from arm and microscope) (if provided) and the remote (for foot switch) cables to the lamp housing. Insert the light guide and plug the mains cable to the sockets on the lamp housing. Installation manual 114 Planmeca Sovereign...
  • Page 119 Make sure that the cables and light guide go through the covers’ opening and bunch them with cable tie as shown on the figure below. Place the power cable to the middle of the cable bunch. Installation manual Planmeca Sovereign 115...
  • Page 120 Carefully attach the covers with four attachment screws. CAUTION Make sure that the cables and light guide are not squeezed. Check the cable routing through the cover’s opening and remove the cable tie. Attach the locking screws. Installation manual 116 Planmeca Sovereign...
  • Page 121 In the base secure the cable with cable ties and make sure that the cables fit under the mains terminal cover. Attach the cover to its position with two attachment screws. The extension of S-video cable is routed out from the base either as concealed wiring or on the floor. Installation manual Planmeca Sovereign 117...
  • Page 122 Remove the cover plug from above the bracket arm attachment screw and unscrew the attachment screw using a 3 mm Allen key. Adjust the rotational friction of the bracket joint with a 3 mm allen key. Tightening the screw increases the friction. Attach the removed covers. Installation manual 118 Planmeca Sovereign...
  • Page 123: Installing The Bracket For Additional Equipment

    This way you can find the correct height of the upper level of the bracket. Turn the bracket assembly slightly and place it to the pylon groove. Slide the bracket assembly to correct height and turn it horizontal position. Tighten the attachment screws using a 5 mm Allen key. Installation manual Planmeca Sovereign 119...
  • Page 124 (1). Attach the tubes to the nipples, green water tube (with additional shielding) to the rear and black air tube to the front of the assembly (2). Tighten the tube attachment nuts using a 12 mm fork spanner (3). Installation manual 120 Planmeca Sovereign...
  • Page 125 If needed, adjust the rotational friction of tray arm by tightening/loosening the adjustment nuts with a 8 mm fork spanner. Adjust the three nuts equally. Tightening the nut increases the friction. Attach the bracket cover to its position with two attachment screws using a 3 mm Allen key. Installation manual Planmeca Sovereign 121...
  • Page 126: Attaching The Pylon Covers

    First slide the pylon side panel upper edge below the bracket (1) and then press the panel to its position (2). Slide the pylon back panel edge below the top bracket and press it to its position. Installation manual 122 Planmeca Sovereign...
  • Page 127: Installing The Assistant Element

    The assistant element rotation limiter screw is attached to the bottom plate during transport (1). Unscrew the limiter screw from the bottom plate and attach it to its position (2). Attach the support strip to the side bars (3). Installation manual Planmeca Sovereign 123...
  • Page 128 Instr. water 3.2/1.6 Green Rinsing water 6/4 Green 01-J7 01-J14 Return water 3.2/1.6 Red 01-J14 Ground cable 01-J7 Suction Air 6/4 Black Rinsing water6/4 Green Instr. water 3.2/1.6 Green Return water 3.2/1.6 Red Ground cable Installation manual 124 Planmeca Sovereign...
  • Page 129 Check that the ground clip is correctly positioned (1) and slide the grounding collar over the clip (2). Secure the Selv cable connected to connector J2 as shown on the figure below. OP-Arm SELV Cable Assistant-Arm SELV Cable Installation manual Planmeca Sovereign 125...
  • Page 130 INSTALLING THE ASSISTANT ELEMENT Lift the assistant element doors to their positions. Installation manual 126 Planmeca Sovereign...
  • Page 131 Attach the assistant element top cover. Loosen the two attachment screws (1). Place the cover to its position and adjust the position so that the gap between the covers is even (2). Tighten the adjustment screws (1). Installation manual Planmeca Sovereign 127...
  • Page 132: Attaching The Suction Tubes

    If the assistant element is equipped with a 3 position holder, a syringe, scaler or LED polymerization light can be attached to the assistant element. If the assistant element is equipped with instrument multiplexer and with 5 position holder, you can attach also rotating instruments to assistant element. Installation manual 128 Planmeca Sovereign...
  • Page 133: Installing The Patient Chair

    Remove the attachment plate from the block of wood and support the frame with the attachment plate. Detach the wheels from the block of wood and attach the wheels to the blocks as shown on the figure below. Installation manual Planmeca Sovereign 129...
  • Page 134 Slide the patient chair into the lifting adapter of the cuspidor. Place the chair in a way that the patient chair supports that are attached to the lifting adapter go to the grooves of the chair adapter. Installation manual 130 Planmeca Sovereign...
  • Page 135 Secure the chair into position with the four attachment screws (shorter screws (1) farther from cuspidor, longer screws (2) closer). Connect the chair cable connector to the cable coming from the cuspidor (3). Attach the grounding cables to the grounding point located on the chair support plate (4). Installation manual Planmeca Sovereign 131...
  • Page 136 INSTALLING THE PATIENT CHAIR Unscrew the wooden frame attachment screws using a 6 mm Allen key (4) and remove the frame (5). Installation manual 132 Planmeca Sovereign...
  • Page 137 (1). Attach the legrest to the seat casting with three attachment screws using a 6 mm Allen key (2). Slide the legrest support to other side of the legrest (3) and secure it with one attachment screw using a 3 mm Allen key (4). Installation manual Planmeca Sovereign 133...
  • Page 138 Attach the legrest upholstery to its position and lift the legrest up. Adjust the legrest cable positions so that the cables do not tighten in upper position. Secure the cables with cable ties (1). Installation manual 134 Planmeca Sovereign...
  • Page 139: Attaching The Seat Bottom Plate

    INSTALLING THE PATIENT CHAIR 12.2 Attaching the seat bottom plate Attach the seat bottom plate (1) to its position with four attachment screws and washers (2). Installation manual Planmeca Sovereign 135...
  • Page 140: Attaching The Upholsteries

    12.3.2 Backrest upholsteries The backrest upholsteries are attached to the backrest casting with attachment pins. Press the backrest upholsteries to their positions so that the attachment pins go into the openings on the upholstery. Attachment pin Installation manual 136 Planmeca Sovereign...
  • Page 141 Slide the legrest upholstery downwards so that the attachment pins go into the grooves of the upholstery’s backplate (1). Attach the lower edge of the backrest upholstery by pressing it towards the legrest (2). Installation manual Planmeca Sovereign 137...
  • Page 142 (1) Pull the edge of the seat upholstery towards the seat (2). Press the legrest down and tighten the two attachment screws located on the front edge of the seat casting. Installation manual 138 Planmeca Sovereign...
  • Page 143: Attaching The Armrests

    12.4 Attaching the armrests If the armrest opening cover is attached to the seat, remove it with a e.g. flat-head screwdriver. Push the armrest firmly to the seat openings until it locks to its position. Installation manual Planmeca Sovereign 139...
  • Page 144: Installing The Side Delivery Arm

    CAUTION MUST BE EXERCISED WHEN CUTTING THE CABLE TIE. Press the arm down with your whole weight when cutting the cable tie (see figure below). Then slowly allow the arm to rise up. Turn the level arm 180 degrees. Installation manual 140 Planmeca Sovereign...
  • Page 145 INSTALLING THE SIDE DELIVERY ARM Place the silicone cover between the arm attachment bracket and base (1) and attach the arm to the base with four attachment screws (2) using a 8 mm Allen key. Tighten the screws firmly. Installation manual Planmeca Sovereign 141...
  • Page 146 INSTALLING THE SIDE DELIVERY ARM Cut the cable ties that attach the support cardboard to the arm and slide the support away from the arm. Installation manual 142 Planmeca Sovereign...
  • Page 147 OP delivery arm” on page 65. Connect the side delivery arm potentiometer cable to the connector P5 on the Mamco PCB located on the cuspidor electronic box. Attach the arm grounding lead to the grounding point located on the cuspidor. Installation manual Planmeca Sovereign 143...
  • Page 148 INSTALLING THE SIDE DELIVERY ARM Push the limiter o-ring and plastic bushing downwards onto the arm axle. Rotate the locking nut down. Installation manual 144 Planmeca Sovereign...
  • Page 149 INSTALLING THE SIDE DELIVERY ARM Remove the level arm lower joint cover. Attach the rotation limiter to its position (see figure below) using a 6 mm Allen key. Installation manual Planmeca Sovereign 145...
  • Page 150 Attach the cover back to its position and secure the cables and tubes with two cable ties. Attach the instrument console to the console bracket using the support plate. Tighten the attachment screws using a 3 mm Allen key. Installation manual 146 Planmeca Sovereign...
  • Page 151 Attach the instrument console cover. Attach the cuspidor and base covers. Attach the arm and joint cover grounding leads to the base grounding points and press the base joint cover to its position. Installation manual Planmeca Sovereign 147...
  • Page 152 Attach the level arm lower joint cover to its position. Pull first the upper part into the arm, then place the lower part to its position. Attach the console arm cover to its position with one attachment screw using a 3 mm Allen key. Installation manual 148 Planmeca Sovereign...
  • Page 153 Refer to section 18.2 “Adjusting the side delivery arm” on page 172. Attach the level arm cover grounding lead to the arm grounding point. Press the level arm to its position. Installation manual Planmeca Sovereign 149...
  • Page 154: Connecting The Foot Control

    Route the foot control cable through the cuspidor bottom plate opening to inside the cuspidor. Open the cuspidor back door. Connect the foot control cable to the connector located at the right side of the electronics control box by pushing it into position. Shut the cuspidor back door. Installation manual 150 Planmeca Sovereign...
  • Page 155: Attaching The Covers

    15.1 Cuspidor bottom plate Slide the cuspidor bottom plate to its position from behind the cuspidor as shown on the figure below (1). Secure the bottom plate to its position with two attachment screws (2). Installation manual Planmeca Sovereign 151...
  • Page 156: Cuspidor Side And Front Covers

    ATTACHING THE COVERS 15.2 Cuspidor side and front covers Attach the cuspidor side and front panels as follows. Push the side panel to its position (1). Attach the two attachment screws (2) of the panel. Installation manual 152 Planmeca Sovereign...
  • Page 157 ATTACHING THE COVERS Push the lower edge of the front panel to the attachment clips (1) and attach the panel upper edge to locking piece (2). Installation manual Planmeca Sovereign 153...
  • Page 158 ATTACHING THE COVERS Attach the cuspidor door to its position. Adjust the hinge positions (1, 2 and 3) so that the door closes properly. Installation manual 154 Planmeca Sovereign...
  • Page 159 ATTACHING THE COVERS Attach the pylon collar (2) to its position with the attachment plate (1) and two screws (2). Attach the cover plate (4). Installation manual Planmeca Sovereign 155...
  • Page 160: Base Cover

    Push the base cover to its position (1). The base cover attachment mechanism is located inside the cuspidor. Rotate the mechanism screw using a 6 mm Allen key to move the mechanism arm and to lock the cover attachment (2). Installation manual 156 Planmeca Sovereign...
  • Page 161: Base Footing Strips

    ATTACHING THE COVERS 15.4 Base footing strips Pull the locking pieces (1) outwards and place the footing strips under them (2). Release the locking pieces. Installation manual Planmeca Sovereign 157...
  • Page 162: Chair Adapter Cover

    ATTACHING THE COVERS 15.5 Chair adapter cover Carefully bend the chair adapter cover (1) and push it to its position. Installation manual 158 Planmeca Sovereign...
  • Page 163: Installing The Instruments

    When you use the holders you can easily shorten or lengthen the hoses, thus move the instrument console to different positions without the hoses hindering the console movements. Secure the locking bar Installation manual Planmeca Sovereign 159...
  • Page 164: Final Adjustments

    The balance of the OP delivery arm should be adjusted when the arm is carrying a nor- mal load, i.e. all the instruments, tray load et.c. Before adjusting the tension of the gas spring remove the lifting friction of the balanced arm with a screwdriver. Loosening the screws decreases the friction. Installation manual 160 Planmeca Sovereign...
  • Page 165 Lift and hold the balanced arm to its upmost position. Adjust the tension of the gas spring by rotating the adjustment screw with a 6 mm Allen key as shown on the figure below. Turning the screw clockwise will increase the tension, i.e, the weight-carrying capacity. Installation manual Planmeca Sovereign 161...
  • Page 166 Adjust the balanced arm so that the arm does not move upwards (or downwards) by itself. Check the whole movement area. Attach the joint cover. Add the lifting friction of the balanced arm slightly so that the arm moves smoothly. Installation manual 162 Planmeca Sovereign...
  • Page 167 Unscrew the attachment screw of the middle joint cover using a 3 mm Allen key. Remove the cover. Adjust the rotational friction of the middle joint with a 3 mm Allen key. Adjust the two screws of the friction block equally. Tightening the screws increases the friction. Installation manual Planmeca Sovereign 163...
  • Page 168 Adjusting the instrument console rotational friction Adjust the rotational friction of the instrument console with the adjustment screw located below the instrument console using a 3 mm Allen key. Tightening the screw increases the fric- tion. Friction adjustment screw Installation manual 164 Planmeca Sovereign...
  • Page 169 Unscrew the four attachment screws of the horizontal arm lower cover using a 3 mm Allen key. Detach the grounding cable from the cover and remove the cover. Remove the cover plug from above the bracket arm attachment screw and unscrew the attachment screw using a 3 mm Allen key. Installation manual Planmeca Sovereign 165...
  • Page 170 FINAL ADJUSTMENTS Adjust the rotational friction of the bracket joint with a 3 mm allen key. Tightening the screw increases the friction. Attach the removed covers. Installation manual 166 Planmeca Sovereign...
  • Page 171 Adjusting rotational friction of the rotating OP delivery arm Remove the joint back cover. Adjust the rotational friction of the OP delivery arm with the adjustment screw located inside the joint using a 3 mm Allen key. Tightening the screw increases the friction Installation manual Planmeca Sovereign 167...
  • Page 172: Adjusting The Side Delivery Arm

    Remove the level arm cover. Before adjusting the tension of the gas spring remove the lifting friction of the balanced arm with a 4 mm Allen key. Loosening the screws decreases the friction. Installation manual 168 Planmeca Sovereign...
  • Page 173 6 mm Allen key as shown on the figure below. Turning the screw clockwise will increase the tension, i.e, the weight-carrying capacity. Add the lifting friction of the level arm slightly so that the arm moves smoothly. Installation manual Planmeca Sovereign 169...
  • Page 174 Adjusting friction of the level arm lower joint Remove the joint cover. Adjust the rotational friction of the joint with a 4 mm Allen key. Adjust the two screws of the friction block equally. Tightening the screws increases the friction. Installation manual 170 Planmeca Sovereign...
  • Page 175 Adjusting the rotational friction of the base joint Remove the joint cover. Adjust the rotational friction of the joint with a 4 mm Allen key. Adjust the two screws of the friction block equally. Tightening the screws increases the friction. Installation manual Planmeca Sovereign 171...
  • Page 176: Assistant Element Adjustments

    FINAL ADJUSTMENTS 17.3 Assistant element adjustments Assistant element arm Adjust the rotational friction of the assistant element arm with a 3 mm Allen key. Tightening the screw increases the friction. Installation manual 172 Planmeca Sovereign...
  • Page 177 FINAL ADJUSTMENTS Assistant element Adjust the rotational friction of the assistant element with a 3 mm Allen key. Tightening the screw increases the friction. Installation manual Planmeca Sovereign 173...
  • Page 178 Suction holder arm The angle of the suction holder arm can be adjusted so that the suction holder is straight. Loosen the adjustment screw (1) and adjust the arm by rotating it (2). Tighten the screw. Installation manual 174 Planmeca Sovereign...
  • Page 179 Adjust the rotational friction of the suction holder arm with a 3 mm Allen key. Note, that there are two screws: one is fully tighten and cannot be rotated and the other is for the friction adjustment. Installation manual Planmeca Sovereign 175...
  • Page 180 Unscrew the four attachment screws of the holder cover and remove the cover. Adjust the rotational friction of the suction holder with the adjustment screw located behind the holder. Tightening the screw increases the friction. Installation manual 176 Planmeca Sovereign...
  • Page 181: Connections

    CONNECTIONS 18 CONNECTIONS Installation manual Planmeca Sovereign 177...
  • Page 182 CONNECTIONS Installation manual 178 Planmeca Sovereign...
  • Page 184 Planmeca Oy | Sorvaajankatu 7 | 00880 Helsinki | Finland tel. +358 20 7795 500 | fax +358 20 7795 555 | sales@planmeca.com | www.planmeca.com...

Table of Contents