2 Pre-installation requirements
Changes to the network
2.6 Patient area
10
Planmeca Compact i3
Changes to the network could introduce risks requiring additional analysis.
Connection to networks including other equipment could result in previously
unidentified risks to patients, operators or third parties. The user
organisation is responsible for identifying, analysing, evaluating and
controlling these risks.
Changes to the IT-NETWORK include:
•
Changes in network configuration
•
Connection of additional items
•
Disconnection of items
•
Update of equipment
•
Upgrade of equipment
The patient area is 1.5 m (59.1") in each direction from the dental unit.
CAUTION
Use only Planmeca specified devices inside the patient area.
CAUTION
The floor of the patient area must be dry.
NOTE
Connect only Planmeca specified devices to the dental unit.
NOTE
The external PC must be protectively earthed and IEC 60950 -approved (CE
marked).
NOTE
The monitor must fulfil IEC 60601-1 ed.3 requirements.