At this point, the interface will prompt whether to stop the running of the Func-
tion, you can press the Left/Right key to select Stop, then the running will be
stopped, and the Function menu item will be displayed, you can re-enter the
Function menu for editing; if Reset is selected, it means to stop the present run-
ning and return to the main interface to wait for the next trigger operation; if
Pause is selected, it means that the present running is suspended, and you can
continue to run by selecting Resume.
4.6.1.1 ARB Function
This series of instruments comes standard with ARB (arbitrary waveform) func-
tion, which allows the output of user-defined arbitrary complex voltage or current
waveforms. It can be operated in the following ways:
•
Import via U disk
The instrument supports importing the edited .csv format file (the template
can be downloaded from ITECH official website or contact ITECH to obtain)
through the USB interface on the front panel, and generating voltage or cur-
rent values after parsing, and then outputting complex voltage or current
waveforms according to the time width specified by the user. The user can
edit a .csv file containing up to 10 million voltage or current points and import
it into the instrument for running to realize arbitrary waveform output or work-
ing condition simulation. The instrument has the ability to quickly parse data.
A .csv file of dozens of megabytes can be imported and parsed within 5S,
making testing more convenient and efficient.
This instrument supports importing the .csv files corresponding to the follow-
ing waveforms:
– CDWELL: Constant dwell arbitrary waveform
– List: User-defined waveform
– Sine: Sine wave
– Sweep: Sweep wave
The steps are as follows:
1. Edit the .csv file corresponding to the template on the PC and save it.
2. Save the edited file in the root directory of the U disk.
3. Insert the U disk into the front panel of the instrument.
4. Press the composite keys [Shift]+[I-set] (Function) on the front panel to
enter the function menu.
5. Select the 1. LIST: Off and press [Enter].
6. Press left/right key to select Open and press [Enter].
Copyright © Itech Electronic Co., Ltd.
Power Supply Function