Table 2 shows the connector order and terminal designation.
Make sure no power is supplied to the instrument.

Strip the insulation on each wire back ¼".

Insert the wire ends through the uppermost opening below

the terminal screw and make sure that the stripped wire
ends are in the terminal, tighten the connector screws.
Figure 9. Terminal strips for analog and relay outputs
RELAY contacts are rated at: 250 V, 2 A AC 500 W

(non-inductive load).
All inputs (except IN 6 & 7) must be connected to

POTENTIAL FREE contacts.
Figure 10. Communication ports
Orion 2210AM Ammonia Analyzer
Installation Environmental Considerations
|
13