Table 3. Error Messages.
Messages
Login as Technician or
Operator
Password mismatch,
Enter to continue
Measurement is in
progress
User programmable
measuring frequency
User programmable
calibration frequency
Two or more inputs are
of same type
Error in date, press F1 to
continue
Error in hours, press F1 to
continue
More than 6 users are not
allowed, delete a user to
add further
No records to delete.
Are you sure you want
to delete all analysis
results?
Are you sure you want
to delete all calibration
results?
No records to display
No records to print
Only one record to display
Problem with memory
allocation. Insufficient
memory in the system
or system is busy.
Orion 2210AM Ammonia Analyzer
Cause
User has not logged in, and trying to
access restricted operations.
User is logged in, but does not have
adequate privileges to access the
restricted operations.
Incorrect password has been entered.
Trying to perform measurement when it is
already under way.
Editing option of analysis frequency is
selected.
If same type of solution is assigned to two
or more inputs.
If same type of solution is assigned to two
or more input.
Date entered is invalid.
Hours entered is invalid.
Technician trying to add more than
6 users.
User is trying to delete Calibration/
Analysis records, when there are no
records stored.
All analysis results? User is trying to delete
all analysis records. (At least one analysis
record is available in the database.)
User is trying to delete all calibration
records. (At least one calibration
record is available in the database.)
User is trying to display records, and no
records are available in the selected data
range.
User is trying to print records, and no
records are available in the selected data
range.
User is trying to view graph when only one
analysis record is available in the
database.
Do not try to repeatedly view the graph.
Solution
User has to login either as a Technician or
Operator to access the restricted
operations.
User is not authorized to access the
restricted operations.
Press "Enter" to cancel the wrong entry and
enable re-entry of correct password, If
password is lost consult factory for back
door access code.
Wait for measurement to complete.
Edit the analysis frequency.
Edit the calibration frequency.
Reassign solution to inputs.
Press F1 and re-enter valid date.
Press F1 and re-enter valid hours.
More than 6 users are not permitted at any
time. To add a new user after 6, delete an
existing user.
None.
None. Deleted records are not recoverable.
None. Deleted records are not recoverable.
None.
None.
None.
Try again after some duration or after
performing one more measurement cycle.