General Product Description - Thermo Scientific 2210AM Manual

Ammonia analyzer
Table of Contents

Advertisement

4. Prime the tubing. (See "Installing Peristaltic Pump Head
Tubing" section for more details).
a.
Press Menu.
b. Select Services.
c. Select Prime Fluidics.
d. Select Prime All.
Note: This procedure will run automatically and
take approximately 10 minutes per cycle.
5. Refer Startup and Operation Section for more details.
Once the above four steps have been completed,
(4. a, b, c, d) press Stop, Select Run. Allow the unit to
operate overnight or for at least 12 hours, to become
stabilized in the local environment.
6. For additional information on Orion 2210AM Ammonia
Process Analyzer Operation, refer Startup and Operation.
Additional details can be found in the following sections:
Instrument Operation
Analyzer Hardware Configuration and Calibration
Calibration of the Instrument for Measurement Analysis
Menu Overview
Terms Definition
Maintenance Schedule
Figure 1. Reagent bottle numbers in proper order.
Orion 2210AM Ammonia Analyzer
General Product
Description
This user manual provides the information necessary to obtain
the maximum benefits from the Orion 2210AM Ammonia
Analyzer. All operating instructions, functional descriptions,
illustrations, and other relevant information are contained in
this manual. Descriptions of the system hardware and
user-interface software (the control panel) are included, as an
overview of the analytical techniques employed in the
analyzer. Step-by-step instructions guide you through all
operating procedures and field programming.
The analyzer is designed for easy operation. Each analyzer in
this family runs in a fully automated mode for routine,
day-to-day water analysis.
The need for operator interaction is minimal. Accordingly,
the information required by the routine operator of this
analyzer is modest, and that information is provided in
Startup and Operation section, of this manual.
Important standard analyzer capabilities include:
Sampling the process stream automatically.

Calculating concentration of your sample and reporting it in

units of concentration.
Recalibrating automatically at preprogrammed intervals to

compensate for any changes in captured sample volume,
pump flow rates and reagent composition.
The analyzer brings colorimetric laboratory precision to the
real world of unattended process analyzing. It is designed for
long life under continuous, rigorous use.
The Orion 2200 series analyzers are designed for
accommodating several applications. Each application
requires a specific configuration in hardware and software that
is set at the factory. The product as supplied will have
a configuration sheet detailing the setup and the detailed tests
and validation that were conducted at the factory. This
analyzer model is intended to use for water applications
which include surface and waste water analysis,
suspended solids must be less than 60 microns.
Introduction
|
5

Advertisement

Table of Contents
loading

Table of Contents