Thermo Scientific 2210AM Manual page 39

Ammonia analyzer
Table of Contents

Advertisement

Terms
CALIBRATION PARAMETERS
Standard
Mode
Frequency
Start Time
Slope
Intercept
Calibration Tolerance
VALIDATION PARAMETERS
Validation Source
Number of Repeats
ANALYSIS PARAMETERS
Mode
Measuring Frequency
Measuring Units
Analysis Label
Cell Wash Flow Rate
Reagent1 and 2 Flow Rate
34 |
Terms Definition
Description
The concentration of the standard used
for calibration.
Mode represents the calibration type.
The time duration between two
calibration cycles.
Calibrations start time, except for manual
and editable frequency.
The CURRENT Slope value represents the
coefficient that is multiplied by the
absorbance to calculate the
concentration of the sample +/- Intercept
value.
Reference and current numbers based on
the base line calibration cycle.
Repeat tolerance tested during auto
calibration cycles to accept or reject
automatic calibrations.
Fluid source for validation.
Number of times that the validation is to
be performed.
Instrument RESTART mode.
In the RUN mode, the frequency is the
time duration between two analysis
cycles.
Units for the analyzed sample.
The rate at which the sample is pumped to
wash the cell.
The rate at which the reagent is pumped
to wash the cell.
Possible Values, Options
0 to XXXXX (XXXXX Value depends on
precision selected in Setup
parameters
result Decimal Point
Single-Point, Base-Line
Every 6 Hr, 12 Hr, 24 Hr, Sun, Mon, Tue,
Wed, Thu, Fri, Sat, Manual and Editable
0 Hr to 99.99 Hr
Depends on Time format in Setup
Time Menu
Default value of -0.5 (depends on range
selected)
Default value of ±5%
Sample, Standard, Grab Sample, Not
Assigned
0 to 8
Auto Restart, Manual Restart
15, 30, 45 minutes, 1 hr, 0.20 hr are
selectable and editable minutes range
00.00 to 99.99 Hr
Six characters long. Alphabets, numerals
and special symbols are allowed.
Six characters long. Alphabets, Numerals
and +, - are allowed.
Depends on the inner diameter of the
pump tubing. It is derived from the
calibration of pumps.
Depends on the inner diameter of the
pump tubing. It is derived from the
calibration of pumps.
Orion 2210AM Ammonia Analyzer
Display