Thermo Scientific 2210AM Manual page 13

Ammonia analyzer
Table of Contents

Advertisement

1. Connect sample to the analyzer and ensure that it is
flowing through the sample chamber by observing the
flow from the sample return line.
2. Turn the power ON to the instrument. After the booting
sequence (this may take several minutes), press STOP
and then MENU key.
3. Select SERVICES MENU option and then
PRIME FLUIDICS
The instrument will go through a priming sequence of all the
available solution to the analyzer. Follow the startup
procedure described in Startup and Operation Section. You
may repeat the AUTO PRIME sequence as necessary to
purge out any trapped bubbles in the lines.
Reagent Consumption
The ammonia method requires two reagents (for extending
the range). The reagent usage depends on the frequency of
analysis and the range selected.
Based on 1 analyses per 2 hours, the reagents should last
approximately 45 days.
8 |
Introduction
AUTO PRIME
ENTER.
Orion 2210AM Ammonia Analyzer