Thermo Scientific 2210AM Manual page 19

Ammonia analyzer
Table of Contents

Advertisement

Table 2. Main terminal strip pin assignments.
Terminal
Designation
Number
1
IN1
2
GND
3
IN2
4
GND
5
IN3
6
GND
7
IN4
8
GND
9
IN5
10
GND
11
IN6
12
RTN
13
IN7
14
RTN
15
+24 V
16
GND
17
Earth Ground
18
GND
19
OUT1
20
Earth Ground
21
GND
22
OUT2
23
NC1
24
NO1
25
C1
26
NC2
27
NO2
28
C2
29
NC3
30
NO3
31
C3
32
NC4
33
NO4
34
C4
14 |
Installation Environmental Considerations
Function
Digital input 1 Live (Level sensor 1 live)
Digital input 1 ground (Level sensor 1 ground)
Digital input 2 live (Level sensor 2 live)
Digital input 2 ground (Level sensor 2 ground)
Digital input 3 live (Level sensor 3 live)
Digital input 3 ground (Level sensor 3 ground)
Digital input 4 live (Level sensor 4 live)
Digital input 4 ground (Level sensor 4 ground)
Digital input 5 live (Level sensor 5 live)
Digital input 5 ground (Level sensor 5 ground)
Digital input 6 live (External 1 control)
Digital input 6 return (External 1 control return)
Digital input 7 live (External 2 control)
Digital input 7 return (External 2 control return)
24 V DC live (limited to 20 mA source)
24 V DC ground
System earth ground
Analog output 1 ground
Analog output 1 (900 Ohms Max)
System earth ground
Analog output 2 ground
Analog output 2 (900 Ohms max)
Normally closed contact relay 1
Normally open contact relay 1
Common contact relay 1
Normally closed contact relay 2
Normally open contact relay 2
Common contact relay 2
Normally closed contact relay 3
Normally open contact relay 3
Common contact relay 3
Normally closed contact relay 4
Normally open contact relay 4
Common contact relay 4