Thermo Scientific 2210AM Manual page 41

Ammonia analyzer
Table of Contents

Advertisement

Terms
Reagent Volume
Stir Time
DATE
Format
Current Date
TIME
Current Time
Format
RECOVERY AFTER POWER FAILURE
Power Fail Duration
Power Fail Clean Cycle
ETHERNET PARAMETERS
IP Address
Server IP
Network ID
RS-232 PARAMETERS
Baud Rate
LANGUAGE
Language
REVIEW / RESET ALARMS
IReset All Alarms?
Reset All Faults?
REVIEW CALIBRATION RESULTS
Display
36 |
Terms Definition
Description
The amount of reagent pumped in to the
cell.
The amount of time the fluid is stirred after
pumping the reagent.
Selectable date format.
Present date, validated depending on the
format.
Present time.
Selectable time format.
Recovery duration after power fail.
Performing clean cycle after recovering
from power fail.
Instrument IP address.
The rate at which the data is transferred
through serial port.
Language in the instrument.
Resets all the raised alarms, which are
shown below this option.
Displays the calibration results for the
selected data range.
Displays the calibration results for the
selected data range.
Possible Values, Options
0 to 6.00 mL
0 to 999 seconds
MM/DD/YY, YY/MM/DD, DD/MM/YY
Legible date
Legible time depending on time format
12 hr, 24 hr
0 to 99.99 hr
Enable, Disable
Editable
115200 (default)
English, other languages, consult factory
Reset All Alarms?
None
None
Orion 2210AM Ammonia Analyzer