Terms
Factory Settings
INSTRUMENT DETAILS
Software Version
Model Number
Serial Number
UNIT LABEL CHANGE
Label Name
PRINTING OPTIONS
Print System
Configuration
Print Analysis Parameters
Print Method Parameters
Print Clean Parameters
Online Print
EDIT CALIBRATED VALUES
Pump1, Pump2,
Pump3, Pump4
Optics Dark
Optics Xref
PRIME FLUIDICS
Priming Volume
PUMPS & STIRRER
Pump1, Pump2,
Pump3, Pump4
Stirrer
RELAYS
Relay1, Relay2,
Relay3, Relay4
38 |
Terms Definition
Description
Select this option to enter into factory
setup.
The current running software version of
the instrument.
Model number of the instrument.
Serial number of the instrument.
Change the title of the instrument.
Prints all the system parameters.
Prints only the analysis parameters.
Prints only the method parameters.
Prints only the clean parameters.
Prints the date, time and concentration
after each analysis cycle.
To change the flow rate of the pumps after
their calibration.
The level detected as dark light level.
The level used as reference potential for
light energy level.
The volume of the fluid selected for
priming.
To check the functionality of pumps.
To check the functionality of stirrer.
To check the functionality of relays.
Possible Values, Options
None
None
None
None
Maximum 9 characters A–Z, 0–9, / , - ,
and space are allowed
None
None
None
None
Enabled, Disabled
0.10 to 60.00 mL
0 to 300 mV
0 to 2400.00 mV
0 to 99.9 mL
ON, OFF
ON, OFF
ON, OFF
Orion 2210AM Ammonia Analyzer