Boston Scientific EKOS Control System 4.0 User Manual page 46

Console
Table of Contents

Advertisement

Message
Error
Description
Code
Check the
E302
Ultrasound
connector�
Check the
E320
Ultrasound
connector�
The Infusion
E304
Catheter and
Ultrasonic
Core are not
compatible�
Ensure the
E314
Ultrasonic Core is
fully inserted in
the catheter�
Black (K) ∆E ≤5.0
Possible Cause
Poor connection of the
1�
Disconnect the Ultrasonic Core or leave disconnected�
Ultrasonic Core�
2�
Inspect for fluid or damage of the pins�
3�
Dry the Ultrasonic Core connector, if necessary�
Ultrasonic Core
4�
Reconnect the Ultrasonic Core and ensure the connection is secure�
Connector is wet at the
5�
Press Start to resume ultrasound�
connection site�
6�
If the error persists, disconnect and reconnect the CIC�
In rare circumstances
7�
Press Start to resume ultrasound�
error code E302 may
8�
If the error persists, please call the phone number listed in the For
also be presented due
Support section�
to a software anomaly�
Poor connection of the
1�
Disconnect the Ultrasonic Core or leave disconnected�
Ultrasonic Core�
2�
Inspect for fluid or damage of the pins�
Ultrasonic Core
3�
Dry the Ultrasonic Core connector, if necessary�
Connector is wet at the
4�
Reconnect the Ultrasonic Core and ensure the connection is secure�
connection site�
5�
Press Start to resume ultrasound�
6�
If the error persists press the Power button to turn off the Control Unit�
7�
Let the Control Unit sit for 30 seconds�
8�
Use the Power button to turn on the Control Unit�
9�
If the error persists, please call the phone number listed in the For
Support section�
The Infusion Catheter
1�
Disconnect the Infusion Catheter and Ultrasonic Core connectors�
and Ultrasonic Core
2�
Inspect for fluid or damage to the pins�
connected to the
3�
Dry the connectors, if necessary�
Control Unit are not
4�
Reconnect the Infusion Catheter and Ultrasonic Core and ensure the
compatible (either
connections are secure�
catheter zone length
5�
Press Start to resume ultrasound�
and ultrasound zone
6�
If the error persists press the Power button to turn off the Control Unit�
length or working
length)�
7�
Let the Control Unit sit for 30 seconds�
8�
Use the Power button to turn on the Control Unit�
9�
If the error persists, please call the phone number listed in the For
Support section�
The Ultrasonic Core is
1�
Verify the Ultrasonic Core is fully inserted in the Infusion Catheter�
not fully inserted in the
2�
Disconnect the Ultrasonic Core connectors�
Infusion Catheter while
3�
Inspect for fluid or damage to the pins�
ultrasound is running�
4�
Dry the connectors, if necessary�
5�
Reconnect the Ultrasonic core and ensure the connections are secure�
6�
Press Start to resume ultrasound�
7�
If the error persists, please call the phone number listed in the For
Support section�
Troubleshooting Step
46