Removal and Installation Procedures
Power Supply
RFI Filter
4–36
You need a screwdriver with a T-15 Torx tip (items 1 and 2).
Removal. To remove the power supply, follow these steps: (See Figure 4-17
on page 4-37.)
1. Remove the handle unit, snaps, cosmetic covers, front-trim unit, EMI covers,
MIO board enforcement bracket, hard disk assembly, drive-bay module, and
Multi-Input/Output board.
2. Orient the instrument so that the bottom is on the work surface.
3. Remove the four screws securing the power supply support bracket.
4. Remove the power supply support bracket from the power supply bracket.
5. Remove the six screws securing the power supply bracket to the chassis.
6. Disconnect the cable from J611 on the Power board.
7. Remove the three screws securing the bracket to the power supply.
8. Carefully slide the module to loosen it from the Power board.
9. Lift the power supply up out of the chassis.
10. Lift the bracket up out of the chassis.
Installation. To install, reverse this procedure.
You need a screwdriver with a T-15 Torx tip (items 1 and 2).
Removal. To remove the RFI filter, follow these steps: (See Figure 4-17 on
page 4-37.)
1. Remove the handle unit, snaps, cosmetic covers, and EMI covers.
2. Orient the instrument so that the bottom is on the work surface.
3. Disconnect the cable from J600 on the Power board.
4. Remove the two screws securing the ground leads to the chassis.
5. Remove the two screws securing the RFI filter to the chassis.
6. Pull the RFI filter away from the chassis.
Installation. To install, reverse this procedure.