Parameter
Choices
UNITS
LB
KG
OZ
G
LB/OZ
ON
LB
KG
OFF
OZ
G
LB/OZ
3.2.3

Calibration Menu

See
Section 4.0 on page 19
CONFIG
WZERO
*CAL*
Display and edit
zero calibration
A/D count value
Parameter
Choices
WZERO
WVAL
WSPAN
REZERO
16
Visit our website
FORMAT Menu
Specifies primary units for displayed and printed weight. Values are: LB=pound; KG=kilogram; OZ=ounce;
G=gram; LB/OZ=pounds and ounces.
Alternate Units (ALTUNT Parameter)
Alternate units. Determines which units are displayed when the UNITS button is pressed. The value for the
primary unit (selected on the PRIMAR parameter) is always ON.
NOTE: The LB/OZ setting is not a Legal-for-Trade setting.
Table 3-3. Format Menu Parameters (Continued)
for calibration procedures.
XXXXXXX
FORMAT
CALIBR
WVAL
Display and edit
test weight value
Figure 3-8. Calibration Menu
CALIBR Menu
Level 2 submenus
Display and edit the zero calibration A/D count value.
Display and edit the test weight value.
Display and edit the span calibration A/D count value.
Press ENTER to remove an offset value from the zero and span calibrations.
Use this parameter only after WZERO and WSPAN have been set. See
tion about using this parameter.
Table 3-4. Calibration Menu Parameters
www.RiceLake.com
Description
SERIAL
XXXXXXX
DIG IN
DEFLT
WSPAN
REZERO
*CAL*
*
Display and edit
Press Enter to
span calibration
remove offset from
A/D count value
zero and span
calibrations
Description
VERS
CAL*
Section 4.0 on page 19
for more informa-