IQ plus 210 Installation Manual
5.0

EDP Commands

The IQ plus 210 indicator can be controlled by a personal computer or remote keyboard connected to the indicator serial port.
Control is provided by a set of EDP commands that can simulate front panel key press functions, display and change setup
parameters, and perform reporting functions. The EDP port provides the capability to print configuration data or to save that
data to an attached personal computer. This section describes the EDP command set and procedures for saving and
transferring data using the serial port.
5.1

The EDP Command Set

The EDP command set can be divided into five groups: key press commands, reporting commands, the
RESETCONFIGURATION special function command, parameter setting commands, and transmit weight data commands.
When the indicator processes an EDP command, it responds with the message
command was received and has been executed. If the command is unrecognized or cannot be executed, the indicator
responds with
.
??
The following sections list the commands and command syntax used for each of these groups.
5.1.1

Key Press Commands

Key press EDP commands (see Table 5-1) simulate pressing the buttons on the front panel of the indicator.
Command
KZERO
KUNITS
KPRIM
KPRINT
KRIGHTARROW
KENTER
KEXIT
5.1.2

Reporting Commands

Reporting commands (see Table 5-2) send specific information to the EDP port. These commands can be used in both setup
mode and normal mode.
Command
DUMPALL
List all parameter values
VERSION
Write IQ plus 210 software version
P
Write current displayed weight.
ZZ
Write current weight and annunciator status.
RS
Reset the indicator
S
Write one frame of stream format
5.1.3

The RESETCONFIGURATION Command

The RESETCONFIGURATION command can be used to restore all configuration parameters to their default values.
This command is equivalent to using the RESET function on the DEFLT menu.
All load cell calibration settings are lost when the RESETCONFIGURATION command is run.
Note
22
Visit our website
Press the ZERO button
Press the UNITS button
Display primary units
Send demand data to serial port
In setup mode, move right in the menu (press the UNITS button)
In setup mode, press the ENTER (ZERO) button
In setup mode only, exit setup mode
Table 5-1. EDP Key Press Commands
Table 5-2. EDP Reporting Commands
www.RiceLake.com
. The
response verifies that the
OK
OK
Function
Function