Revolution ™ Calibration - Rice Lake IQ plus 210 Installation Manual

Digital revolution scale software
Table of Contents

Advertisement

4.3
Revolution
Calibration
To calibrate the indicator using Revolution, the indicator must be in setup mode with the EDP port connected to a PC running
the Revolution configuration utility. Use the following procedure to calibrate the indicator:
1. Select Calibrate Indicator from the Revolution main menu.
2. On the Weight Calibration display, select the indicator model (IQ+210) and communications port then click OK.
3. Revolution uploads calibration data from the indicator then presents the information in a display like that shown in
Figure
4-3.
4. Enter the Value of Test Weight to be used for span calibration then click OK.
5. The Zero Calibration dialog box prompts you to remove all weight from the scale. Clear the scale and click OK to begin
zero calibration.
NOTE: If your test weights require hooks or chains, place the hooks or chains on the scale for zero calibration.
Note
6. When zero calibration is complete, the Span Calibration dialog box prompts you to place test weights on the scale for
span calibration. Place tests weights on the scale then click OK.
7. The Rezero dialog box prompts you to remove all weight, from the scale (including hooks or chains). Remove the
weights then click OK.
8. When calibration is complete, the New Settings fields of the Weight Calibration display are filled in. Click Exit to save
the new values and return to the Revolution main menu; to restore the previous calibration values, click Restore
Settings.
Figure 4-3. Revolution Calibration Display
© Rice Lake Weighing Systems ● All Rights Reserved
Calibration
21