CAUTION:
This product has an autoranging line voltage input. Be sure that the supply voltage is within
the specified range.
WARNING
CAUTION

Meeting Electrical and Environmental Requirements

Line Settings
The available AC power source must meet the following requirements:
Voltage:
100/115 volts nominal (90-132 volt range)
230/240 volts nominal (198-254 volt range)
Frequency:
for 100/115 volts: 50/60/400 Hz nominal
for 230/240 volts: 50/60 Hz nominal
Power:
200 watts maximum
Verify that the power cable is not damaged and that the power source socket outlet
provides a protective earth contact.
This is a Safety Class 1 Product (provided with a protective earthing ground
incorporated in the power cord). The mains plug shall only be inserted in a
socket outlet provided with a protective earth contact. Any interruption of
the protective conductor inside or outside of the product is likely to make the
product dangerous. Intentional interruption is prohibited.
Always use the three-prong AC power cord supplied with this product. Failure to
ensure adequate earth grounding by not using this cord may cause product
damage.
Environment
This product is designed for use in the following environmental conditions:
indoor use
< 15,000 feet (4,572 meters) altitude
0° to 55° C temperature, unless specified differently
maximum relative humidity 80% for temperatures up to 31° C, decreasing linearly to
50% relative humidity at 40° C
Preparing for Use
Installing the Signal Generator
1-5