Immediate; Internal - HP ESG Series User Manual

Signal generators
Hide thumbs Also See for ESG Series:
Table of Contents

Advertisement

Immediate

Internal

Hardkey and Softkey Reference
AM
SCPI Commands:
[:SOURce]:AM[1]|2:SOURce EXTernal2
[:SOURce]:AM[1]|2:EXTernal2:COUPling DC
[:SOURce]:AM[1]|2:EXTernal2:COUPling?
This softkey is one of the choices in the AM Sweep Trigger menu. Choosing
immediately triggers swept-sine amplitude modulation in a continuous
Immediate
mode. To stop the sweep, choose a different trigger such as
waveform to something other than
Softkey Location: Press
Immediate
Status after Normal Preset: AM sweep trigger is set to Immediate
SCPI Commands:
[:SOURce]:AM[1]|2:INTernal[1]:SWEep:TRIGger IMMediate
[:SOURce]:AM[1]|2:INTernal[1]:SWEep:TRIGger?
This softkey lets you connect an internally-generated, calibrated signal to the AM
modulator. The default state of the internal waveform generator is a 400 Hz
sinewave with 0.1% depth.
The internal source has multiple uses but can be used for only one modulation at a
time. If, for example, you were using the internal source in a frequency
modulation configuration for
also use the internal source, the signal generator would turn off
assign the internal source to your
Softkey Location: Press
SCPI Commands:
[:SOURce]:AM[1]|2:SOURce INT[1]
[:SOURce]:AM[1]|2:SOURce?
5-16
.
Swept-Sine
,
,
AM
AM Waveform
Swept-Sine
, and then you configured
FM Path 1
configuration.
AM Path 2
,
,
AM
AM Source
Internal
, or set the
Ext
,
,
AM Sweep Trigger
to
AM Path 2
and
FM Path 1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Esg-1000aEsg-2000aEsg-3000aEsg-4000a

Table of Contents