Troubleshooting
ESG Series Signal Generator Instrument-Specific Error Messages
Configuration error; Invalid Data Generator
memory configuration.
This indicates that the memory configuration for an option
board does not match the known memory limits of any
supported option board. If this error has occurred as the result
of a customer-installed option, uninstall all options and then
reinstall the correct options. If the error persists, contact the
factory.
Configuration error; Invalid option board
configuration.
This indicates that an invalid combination of option boards has
been configured. If this error has occurred as the result of a
customer-installed option, uninstall all options and then
reinstall the correct options. If the error persists, contact the
factory.
700
State Save Recall Error;Recall aborted. Unable
to recall the state from non-volatile memory.
This indicates that the state file was not readable, so the recall
was aborted. If state file exists, delete explicitly using the
memory catalog.
State Save Recall Error;Recalled state has a
bad checksum. No instrument state change.
This indicates that the state file was corrupt or out-of-date, so
the recall was ignored. If state file exists, delete explicitly using
the memory catalog.
State Save Recall Error;Recall data different
from FW revision. No instrument state change.
Indicates that an attempt was made to recall a state that was
saved with an incompatible version of the instrument firmware.
This typically occurs when a state file is copied from an
instrument with a newer version of firmware to an instrument
with an older version of firmware. Newer versions of
instrument firmware can read older state files.