AM Start Rate

Hardkey and Softkey Reference
AM
Use this softkey to change the starting internal modulation frequency for
swept-sine amplitude modulation. To change the value, press the softkey once.
The current value for the AM start rate is displayed in the active entry area. To
enter a new value, rotate the front panel knob until the desired value is displayed,
use the up and down arrow keys, or enter the value using the numeric keypad and
or
terminator softkey. The range of values allowed is 0.1 Hz to 50
press the
Hz
kHz
kHz. The minimum increment allowed is 0.1 Hz.
This softkey is coupled to the
softkey in the AM Dual-Sine Menu. Any value set for
AM Tone 1 Rate
is reflected in the
AM Start Rate
if you change the value for either the
softkey, the value for
AM Start Rate
Softkey Location: Press
Status after Normal Preset: 400.0 Hz
SCPI Commands:
[:SOURce]:AM[1]|2:INTernal[1]:FREQuency <val><unit>
[:SOURce]:AM[1]|2:INTernal[1]:FREQuency?
5-8
softkey in the initial AM Menu and the
AM Rate
and
AM Rate
AM Tone 1 Rate
softkey or the
AM Rate
is changed to match.
,
,
AM
AM Waveform
Swept-Sine
softkeys. Conversely,
AM Tone 1 Rate
,
AM Start Rate