Download Print this page

Otto Bock Axon-Bus Instructions For Use Manual page 34

Prosthetic system

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 24
6.4 Safety Shutoff
The purpose of the Axon-Bus prosthetic system safety shutoff is to protect the battery; it is
triggered in case of:
Excessively high or low temperature
Overvoltage or undervoltage
Short circuit
After a short circuit, the charging plug must be plugged into the charging receptacle and then
unplugged in order for the electronics to be activated.
7 Use
7.1 Switching the Axon-Bus prosthesis system on and off
INFORMATION
Switching off the Axon-Bus prosthetic system during extended passive pauses (e.g. air or rail
travel, visit to a theatre or cinema, etc.) will make the battery last longer before requiring a
recharge. Only the entire Axon-Bus prosthetic system with all connected components can be
switched off. Individual Axon-Bus components cannot be switched off separately.
7.2 Opening the Axon-Bus gripping component in an emergency
7.3 Changing the Axon-Bus gripping component
CAUTION
Changing Axon-Bus gripping components when switched on
Injury due to faulty control or malfunction of the Axon-Bus prosthetic system.
Power down the Axon-Bus prosthetic system by pressing the button in the charging recept­
acle before changing any Axon-Bus components (e.g. Axon-Bus gripping component).
Attaching the Axon-Bus gripping component to the socket
1) Slide the terminal device on to the socket (rotation anchored in the socket) until it locks in
place audibly.
2) Verify proper attachment by pulling on the terminal device.
34
► 1) Push and hold the charging receptacle button until a confirma­
tion sound is heard (at least 1 second).
→ The prosthesis and prosthetic components are switched on.
Repeat this process to switch the prosthesis and prosthetic compon­
ents off.
This safety function allows the Axon-Bus gripping component to be
opened regardless of the control signals present.
1) With the Axon-Bus prosthetic system switched on, press the char­
ging receptacle button and hold for approximately three seconds
until the Axon-Bus gripping component begins to open.
→ A pulsating beep sounds as it opens.
2) Releasing the button immediately stops the process of opening
the Axon-Bus gripping component and turns the entire Axon-Bus
prosthetic system off.

Advertisement

loading