Download Print this page

Otto Bock Axon-Bus Instructions For Use Manual page 43

Prosthetic system

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 24
Charging recept­
acle
-
-
Battery charger
Error while charging the product
Battery charger
12.2.3 Status signals
Battery charger is connected
Battery charger
Beeper
3 x short
Pulsating
Explanation/correction
An error has occurred in the Axon-Bus prosthetic sys­
tem
Consult prosthetist
Axon-Bus prosthetic system requires service
Consult prosthetist
Explanation/correction
Axon-Bus prosthetic system is not charging
Disconnect and then reconnect the charging plug.
Clean charger contacts.
Charger is overheated or overcooled
Disconnect charging plug and allow charger to cool
down/warm up.
Battery charger or power supply unit is defective
Consult prosthetist
Explanation
Power supply and battery charger operational
Battery is charging
Battery is charging, battery 25% charged
Explanation/correction
Axon-Bus prosthetic system switches itself
off
Charge battery
Opening the Axon-Bus gripping component
in an emergency
System is switched off, system can be used
without restriction after being switched on
again
43

Advertisement

loading