Medical device
Legal manufacturer
Protection against penetration of solid foreign objects with a diameter > 2.5 mm,
protection against water spray
Lot number (PPPP YYYY WW)
PPPP – plant
YYYY – year of manufacture
WW – week of manufacture
Article number
12.2 Operating states/error signals
The product indicates operating states audibly and visually.
12.2.1 Signals for operating states
Battery charger connected/disconnected
Beep signal
1 x short
2 x short
1 x long
Switching the Axon-Bus prosthetic system on/off
Charging recept
acle
/
/
12.2.2 Warnings/error signals
Error during use
Charging recept
acle
42
Event
Battery charger connected
Charging begins
Charging ends
Beeper
2 x short
1 x long
Beeper
1x long
-
Event
Switch on the Axon-Bus prosthetic system
Switch off the Axon-Bus prosthetic system
Explanation/correction
Error in the Axon-Bus prosthetic system
•
Switch the Axon-Bus prosthetic system off/on
•
Open/close the Axon-Bus prosthetic system
•
Consult prosthetist if error persists
Axon-Bus prosthetic system outside operat
ing temperature range
•
Allow the Axon-Bus prosthetic system to
warm up or cool down (Technical data)