Setting Run Temperature

Select a run temperature between –10 and +40°C, in 1°C increments.
NOTE
For runs at temperatures other than ambient, always refrigerate or warm the rotor to the required run
temperature before the run. For low-temperature runs, precool the system by running a precooled rotor
at 2000 rpm at the required temperature for at least 30 minutes.
1
Turn the
• After you release the knob, the display shows the selected temperature.
• After 5 seconds, the display reverts to the actual current chamber temperature.
• When the run begins, the display shows the rotor temperature ±2°C (after a period of
temperature equilibration).
The system controls the chamber temperature to within 2°C of the set temperature.
NOTE
temperature within 1°C of set, perform the temperature calibration procedure provided in the
Appendix.

Setting Acceleration and Deceleration Rates

Table 3.1
and
graphic representation of these settings.
Table 3.1 Acceleration Settings
Maximum Acceleration
(not available with the
JS-24.38 and JS-24.15 rotors)
Full torque is used from 0 rpm
until set speed is reached
Application: pelleting
* During transient conditions, such as acceleration and deceleration, rotor temperature may be outside this range.
J326XP-IM-5AB
knob until the required temperature appears in the display.
TEMP°C
Table 3.2
describe the acceleration and deceleration settings.
Reduced torque is used from 0 to 500 rpm, lasting approximately
2 minutes. (The time varies depending on the mass of the rotor in use.
Larger rotors take longer.). Above 500 rpm, full torque is used until set
speed is reached.
Application: gradients
Entering Run Parameters
Figure 3.1
Slow Acceleration
3
Operation
*
To control
provides a
3-7