Starting A Run - Beckman Coulter Avanti J-26XP Instructions For Use Manual

High-performance centrifuge
Hide thumbs Also See for Avanti J-26XP:
Table of Contents

Advertisement

7
When the rotor speed has stabilized, look through the port in the centrifuge door and turn the
knob until the elutriation chamber in the rotor is synchronized with the strobe.
DELAY
• (The chamber will appear to be motionless.)
• Sample can now be injected into the lines.
• See the rotor instruction manual for detailed instructions.
NOTE
black rubber stoppers in the ports at the left side of the centrifuge door. Failure to seal these openings
will prevent proper operation of the FRS system, which will cause a diagnostic shut down during the
next run.

Starting a Run

1
Check that all parameters are correct and the door is firmly closed.
2
Press
START.
• The display will start showing actual centrifuge values within 5 seconds.
The LED on the
• When set speed is reached, the light stays on continuously until the centrifuge begins
decelerating.
• As the run begins, the system identifies the installed rotor.
• The system verifies that the user-selected rotor and the installed rotor are the same.
• If the system detects that the installed rotor is not the rotor that the user selected with the
ROTOR
maximum permitted speed, the system will reduce the run speed to the maximum allowed
for the installed rotor.
J326XP-IM-5AB
1. Delay Control Knob
At the end of the run, after disassembling the liquid and power lines, be sure to replace the
key flashes during acceleration.
START
knob, and the speed selected by the user is greater than the installed rotor's
DELAY
1
ELUTRIATOR
3
Operation
Entering Run Parameters
3-11