Specifications and Preinstallation Requirements
Specifications
Specifications
Acceleration
Deceleration
a. During transient conditions, such as acceleration and deceleration, rotor temperature may be outside this range. Refer to
applicable rotor manuals for specific rotor operating range information.
b. To reach temperatures above ambient, the centrifuge is dependent on the frictional heat generated inside the chamber
during operation. At low run speeds or low ambient temperatures, the centrifuge may not be able to achieve some higher
temperatures.

Operational Features

Specifications
Door
Rotor Chamber Diameter
Friction Reduction
System (FRS)

Physical Data

Specification
Width
Depth
Height
Weight
Clearances
(for adequate
ventilation)
Surface Finish
1-2
Description
maximum or slow
maximum, slow, or off
Description
6.1-cm (2.4-in.) thick structural foam with steel plate
51.3 cm (20 in.)
190 mm (7.5 in.) Hg
Description
71 cm (28 in.)
• 86 cm (34 in.)
• including air diverter extending from back panel
— 102 cm (40.25 in.)
• with door closed
— 86 cm (34 in.)
• to top of control head
— 116 cm (45.5 in.)
• to top of open door
— 149 cm (58.7 in.)
290 kg (640 lb)
• sides
— 7.6 cm (3 in.)
• back (place air diverter against the wall)
— 16 cm (6.25 in.)
polyester control panel with polycarbonate coating overlay; polyurethane
enamel on door and covers; acrylic baking enamel on other surfaces
J326XP-IM-5AB