Hold the left top edge in place and insert the Allen wrench through the hole.
a. Turn to the right (clockwise) about one-half turn, until you reach a stop.
• The latch engages.
3
Repeat step 2 on the right side
• The second latch engages.
Figure 4.4 Front Panel Retaining Strip
1
1. Retaining Strip

JCF-Z Rotor Identification

The rotor identification system can, under two specific conditions, misidentify the JCF-Z
continuous flow/zonal rotor. Misidentification has occurred if a rotor name other than JCF-Z
appears in the
misidentified, stop the run and follow one of the steps below.
NOTE
Misidentification can be prevented by returning your JCF-Z rotor to the Beckman Coulter factory to
have magnets added to the rotor body. The magnets ensure correct identification. Call your local
Beckman Coulter Field Service representative
JCF-Z rotors manufactured after March, 1997, have factory-installed magnets.
* In the United States, call 1-800-742-2345. Outside the U.S., contact your local Beckman Coulter office.
J326XP-IM-5AB
display during the run, when rotor identification is complete. If the rotor is
ROTOR
*
for more information.
4
Troubleshooting
JCF-Z Rotor Identification
4-9