With the Maxwell
Sample Track software you also can update the firmware through the RS-232 port on the back of
®
the instrument. Firmware can be updated from any computer running Microsoft
Maxwell
Sample Track Software Technical Manual #TM314 for details.
®
11.B. Service
16 Forensic Instrument allows you to print the service history of the instrument. In the Home screen,
The Maxwell
®
select Setup. In the Maxwell 16 Configuration screen, select Service Mode, and enter the Administrative PIN that is
supplied in the Welcome letter to access the Utilities Menu screen.
Figure 42. Utilities Menu screen.
In the Utilities Menu screen, select Service Report to view and print a current service report. The report will list the
report date, instrument serial number, date when last service and preventative maintenance were performed, number
of run cycles the instrument has performed, total number of hours the UV lamp has run and firmware version.
You also can print the service report using the Maxwell
describes how to do this.
The Utilities Menu screen allows the following options:
Run Count shows the number of times the instrument has been run.
UV Bulb Time shows the number of hours the bulb has been used. If you are replacing the bulb (see
Section 8.E.), you may reset the number of hours to zero by selecting the Yes option in the UV bulb screen.
Reset Name allows you to rename the instrument. Use the Run/Stop button to add the new name.
Reset Wizard opens the Reset Wizard screen.
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
www.promega.com
®
Sample Track software. See Technical Manual #TM314, which
Windows
XP or Vista
®
®
TM321 · Revised 4/15