Troubleshooting (continued)
Symptoms
UV lamp errors.
42
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
TM321 · Revised 4/15
Causes and Comments
• After 2950 hours of accumulated lamp time, the following
error message appears.
Figure 40. UV Lamp Time. This message means that the UV
Bulb has reached the limit of its rated life span and should be
replaced. See Section 8.E.
• If you see the error message below, the bulb may not be
present or installed properly (see Figure 10). If the bulb is
installed properly, this may mean that the bulb is not
functioning. It should be replaced. See Section 8.E.
Figure 41. UV lamp error.