7.A. Changing the Hardware
Important. Do not reconfigure the instrument hardware while samples or reagent cartridges are in place on
!
the instrument platform. Clean the instrument thoroughly following the instructions in Section 8 before
removing the existing hardware. Wear gloves when changing the hardware to avoid exposure to any potential
contaminants.
Changing from SEV to LEV Hardware
Select Hardware on the Maxwell
1.
2.
Follow the on-screen instructions. First, open the door.
3.
The Magnetic Assembly will move down to allow you to remove the Magnetic Rod Assembly. Do not push down
on the Magnetic Rod Assembly if the instrument is powered on.
4.
Unscrew the three thumbscrews (boxes in Figure 32) on top of the Magnetic Rod Assembly.
Figure 32. SEV Magnetic Rod Assembly and thumbscrews.
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
www.promega.com