7.A. Changing the Hardware (continued)
5.
Gently lift the SEV Magnetic Rod Assembly, and remove it.
Figure 33. Removing the SEV Magnetic Rod Assembly.
6.
Insert the black LEV Plunger Bar Adaptor onto the bottom of the remaining arm. Make sure that the side
containing the screws faces upward. Several magnets will "click" and hold the LEV Plunger Bar Adaptor in place
once it is attached. The LEV Plunger Bar Adaptor should then be fairly difficult to remove.
Figure 34. Inserting the LEV Plunger Bar Adaptor.
7.
Insert the black LEV Magnetic Rod Assembly in the correct orientation. The screws should be facing toward you.
Press firmly to place the LEV Magnetic Rod Assembly in position. It will fit tightly in place.
34