Promega Maxwell 16 Forensic Technical Manual page 36

Table of Contents

Advertisement

Figure 35. Orientation of the Magnetic Rod Assembly.
8.
Secure the LEV Magnetic Rod Assembly with the three thumbscrews from the SEV Magnetic Rod Assembly, and
hand-tighten.
9.
Close the door as indicated on screen. The screen will indicate that the hardware has been changed. The firmware
indicates a hardware change once the door is closed whether or not you actually completed the physical hardware
change. Confirm that the hardware shown on the screen display matches the physical hardware installed. If the
screen does not match the physical hardware installed, turn the instrument off, install the appropriate hardware
and restart the machine. The hardware indicated on-screen should match the physical hardware installed.
Important. When the Home screen is accessed, it will indicate the new hardware mode. Always confirm that the
!
hardware mode shown matches the physical hardware installed. Running the instrument if the hardware mode
shown does not match the physical hardware installed will damage the instrument.
Changing from LEV to SEV Hardware
1.
Select Hardware on the Maxwell
2.
Follow the on-screen instructions. First, open the door.
3.
The Magnetic Assembly will move down to allow you to remove the Magnetic Rod Assembly. Do not push on the
Magnetic Rod Assembly if the instrument is powered on.
4.
Unscrew the three thumbscrews (boxes shown in Figure 36) on top of the Magnetic Rod Assembly.
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
www.promega.com
Back
16 Configuration screen.
®
Front
35
TM321 · Revised 4/15