In the User Settings screen, select Edit/Add User, press the Run/Stop button and add the user name using the
keyboard.
Note: The user name is limited to 15 characters.
Once the user name is added, press the Run/Stop button.
The Wizard will return to the User Settings screen. A PIN can be added for each user for security. PIN protection
requires the user to enter a PIN to perform a purification run. The PIN is four digits. The list of users and associated
PINs can be modified as described in Section 7.B. Keep a separate list of all users and associated PINs in case a PIN is
forgotten.
An administrative PIN is included in the Welcome Letter, sent with the instrument, which is used to access all PIN-
protected options. If you lose your Administrative PIN, contact Promega Technical Services.
To add a PIN, select Edit/Add PIN in the User Settings screen. A keypad will open. Add the desired PIN, and press
the Run/Stop button. Confirm the PIN, and press the Run/Stop button.
In the User Settings screen, selecting Exit will take you to the User Setup Complete screen, where you can choose
to add more users. Selecting Yes will return you to the Select User screen, where the added user is highlighted. Select
another number, and repeat the process to add another user and PIN. Repeat until all users and PINs are added, then
select the Exit button in the User Setup Complete screen. Select No to end the Wizard. A maximum of 10 users can
be added.
The Wizard will indicate that it is finished and will restart the instrument. Upon restart, the instrument will perform a
self-diagnostic test prior to opening the Home screen. The platform, Plunger Bar and Magnetic Rod Assembly are
moved to check calibration, and the UV light is briefly switched on to ensure the bulb is functioning.
During the diagnostic test the firmware version is displayed on the screen. If you need to update to a new firmware
version, see Section 11.
If you have chosen to run the UV lamp each time the instrument is switched on, the UV lamp will come on and run for
the specified time. You will have the option to cancel this treatment.
!
Important. Do not open the door if the UV lamp is on.
UV lamp will not operate with door open.
The Home screen contains the menu options that allow access to all functions available for the instrument and displays
the hardware mode (SEV or LEV).