7.A. Changing the Hardware (continued)
Figure 36. LEV Magnetic Rod Assembly and thumbscrews.
5.
Gently lift the LEV Magnetic Rod Assembly, and remove it. Pull down on the LEV Plunger Bar Adaptor to release
the magnets that hold it in place.
Note: Pull down at an angle so that the back of the magnet is released first.
6.
Remove the black LEV Plunger Bar Adaptor.
Figure 37. Removing the LEV Plunger Bar Adaptor.
7.
Insert the SEV Magnetic Rod Assembly. The set screws should be facing toward you. Press firmly to place the
Magnetic Rod Assembly in position. It will fit tightly into place.
Figure 38. Installing the SEV Magnetic Rod Assembly.
36
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516
TM321 · Revised 4/15
www.promega.com