Chapter 6 Operations and Measurements
Y: It is a pair of horizontal solid (Cursor A)/dotted (Cursor B) lines, used for
measuring voltage parameters. The measurement results include AY, BY,
and △Y.
2. Select Measurement Source
Press Source, then rotate the multifunction knob
source (CH1-CH4). Press down the knob to select it. You can also press Source
continuously or enable the touch screen to select it. The available source
channels are None, CH1-CH4, LA, and Math1-Math4.
Note:
Only channels currently enabled can be selected.
When "LA" is selected under Source, Select is grayed out and disabled. By
default, "X" is selected under Select.
3. Adjust Cursor Position
1) When "X" is selected under Select, you can adjust the position of X cursor.
Press AX, then rotate the multifunction knob
horizontal position of Cursor A (X cursor). Its adjustable range is
limited within the screen.
Enable the touch screen and tap AX, then the numeric keypad is
displayed. Set the horizontal position of Cursor A (X cursor). The
horizontal axis indicates time, and the unit of its setting value is the
same as that of the horizontal axis.
Press BX, then rotate the multifunction knob
horizontal position of Cursor B (X cursor). Its adjustable range is
limited within the screen.
Enable the touch screen and tap BX, then the numeric keypad is
displayed. Set the horizontal position of Cursor B (X cursor). The
horizontal axis indicates time, and the unit of its setting value is the
same as that of the horizontal axis.
Press AX BX, then rotate the multifunction knob
horizontal position of Cursor A and Cursor B (X cursors) simultaneously.
Its adjustable range is limited within the screen. The horizontal spacing
between Cursor A and Cursor B (X cursors) remains unchanged.
2) When "Y" is selected under Select, you can adjust the position of Y cursor.
Press AY, then rotate the multifunction knob
position of Cursor A (Y cursor).
Enable the touch screen and tap AY, then the numeric keypad is
displayed. Set the vertical position of Cursor A (Y cursor). The vertical
axis indicates voltage, and the unit of its setting value is the same as
that of the vertical axis.
MSO8000 User Guide
to select the desired
to adjust the
to adjust the
to adjust the
to adjust the vertical
RIGOL