Remote Control Via Lan - Rigol MSO8064 User Manual

Digital oscilloscope
Hide thumbs Also See for MSO8064:
Table of Contents

Advertisement

RIGOL

Remote Control via LAN

1. Connect the device
Use the network cable to connect the oscilloscope to your local area network
(LAN).
2. Configure network parameters
Configure the network parameters of the oscilloscope according to "LAN
Configuration".
3. Search for the device resource
Start up Ultra Sigma and click
figure below. Click
resource currently connected to LAN. The resource found is displayed at the
right of the panel. Click
address of the instrument manually into the text field under "Manual Input LAN
Instrument IP", then click
to add the instrument to the LAN instrument resource list in the right
section; if the instrument fails the test, please check whether the IP address that
you input is correct, or use the auto search method to add the instrument
resource.
4. View the device resource
The resources found will appear under the "RIGOL Online Resource" directory.
For example, MSO8104 (TCPIP::172.16.3.14::INSTR).
5. Control the instrument remotely
Right-click the resource name "MSO8104 (TCPIP::172.16.3.14::INSTR)" and
select "SCPI Panel Control" to turn on the remote command control panel,
through which you can send commands and read data.
6. Load LXI webpage
As this oscilloscope conforms to LXI CORE 2011 DEVICE standards, you can load
LXI webpage through Ultra Sigma (right-click the instrument resource name and
21-6
to open the panel as shown in the
and the software searches for the instrument
to add it. Besides, you can input the IP
. If the instrument passes the test, click
Chapter 21 Remote Control
MSO8000 User Guide