Rigol MSO8064 User Manual page 128

Digital oscilloscope
Hide thumbs Also See for MSO8064:
Table of Contents

Advertisement

RIGOL
Press Bytes, then rotate the multifunction knob
numeric keypad to set the length of the data. Its range is from 1 to 5.
Press More → Addr Bits, and rotate the multifunction knob
the desired address bits. Then press down the knob to select it. You can
also press Addr Bits continuously or enable the touch screen to select it.
The available address bits are "7 Bits", "8 Bits", and "10 Bits".
A & D: the oscilloscope searches for the specified address and data at the same
time, then triggers when both the address and data meet the conditions. After
this trigger condition is selected, you need to set Bit X, Bytes, Addr Bits,
Address, and Direction menu items. For setting methods, refer to descriptions
in "Address" and "Data" trigger conditions.
Trigger Mode:
In the trigger control area (Trigger) on the front panel, press Mode to quickly switch
the current trigger mode. For details, refer to descriptions in "Trigger Mode".
Trigger Parameter Setting:
Set the trigger parameter (noise rejection) under this trigger type. For details, refer
to descriptions in "Noise Rejection".
Trigger LEVEL/Threshold Level:
Rotate the Trigger
to "Trigger LEVEL/Threshold Level". The current trigger level/threshold level
value is displayed at the upper-right corner of the screen.
5-38
LEVEL knob to adjust the trigger level/threshold level. Refer
Chapter 5 To Trigger the Oscilloscope
or use the pop-up
MSO8000 User Guide
to select