Rigol MSO8064 User Manual page 247

Digital oscilloscope
Hide thumbs Also See for MSO8064:
Table of Contents

Advertisement

Chapter 12 Protocol Decoding
information will be displayed. For example, when "none" is chosen for the
parity on the transmitter and "odd" is chosen for the parity on the decoder,
the following parity error information
value of the horizontal time base) occurs. Wherein, the number of 1s in the
8-digit figure (01001111) is an odd number, and the parity bit should be 0.
However, the parity bit detected on the Tx is 1; thus a parity error occurs.
EOF Error
This error is generated when the EOF condition is not met. For example, if
the Stop Bit is set to 1.5, but the actual stop bits is less than 1.5 bits, then
the red error report information
of the horizontal time base) is displayed in the following figure.
Parity Error and EOF Error
When both the parity error and the EOF error are detected, the error report
information (its display form is related to the value of the horizontal time
base), as shown in the following figure, will be displayed.
MSO8000 User Guide
(its display form is related to the
(its display form is related to the value
RIGOL
12-13

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mso8000 seriesMso8104Mso8204

Table of Contents