Rigol MSO8064 User Manual page 86

Digital oscilloscope
Hide thumbs Also See for MSO8064:
Table of Contents

Advertisement

RIGOL
Note:
The sample rate is displayed both in the status bar at the upper part of the
screen and in the CH SampleRate menu. To indirectly modify the sample
rate, rotate the Horizontal
base or modify the memory depth. In addition, if the triggering source selects
the analog channel, it will cause the channel mode to change, affecting the
sampling rate.
The impact of low sample rate on the waveform:
1. Waveform Distortion: when the sample rate is too low, some waveform
details are lost, and the sample waveform displayed is rather different from the
actual waveform of the signal.
2. Waveform Aliasing: when the sample rate is twice lower than the actual
signal frequency (Nyquist Frequency), the frequency of the waveform rebuilt from
the sample data is smaller than the actual signal frequency.
3. Waveform Leakage: when the sample rate is too low, the waveform rebuilt
from the sample data does not reflect all the actual signal information.
4-8
Chapter 4 To Set the Sample System
SCALE knob to adjust the horizontal time
MSO8000 User Guide