DC Power
Battery Power
Battery Charge Status
Battery In Use
Revision A
Equipment Overview: Indicators
Color
Green
The monitor is connected to an electrical outlet.
None
The monitor is not connected to an electrical outlet.
Color
Yellow
The monitor is powered by an internal battery.
The monitor is powered by a TRAM-RAC housing.
None
The monitor is powered by the external power supply.
There are two battery change status indicators; the left indicator
displays the status of battery A, the right indicator displays the status of
battery B.
Color
Green
The respective internal battery is fully charged.
Yellow
The respective internal battery is being charged. If both batteries
are present and require charging, then both indicators illuminate,
even though they charge sequentially.
None
The respective internal battery is not installed.
The respective battery has failed.
The monitor is powered by the external power supply.
The monitor is powered by the respective internal battery.
There are two battery-in-use indicators; one inside each battery door.
Color
Green
The monitor is powered by the internal battery. If the indicators
for both batteries are illuminated, the monitor is in a very low
battery condition and must run off both batteries to sustain power.
None
The monitor is powered by the external power supply.
The monitor is powered by both batteries simultaneously.
Transport Pro
2012659-042
Description
Description
Description
Description
2-9