TRAM

Revision A
Disassembly: FRU Replacement
WARNING
DROP HAZARD — Install the TRAM chute mounting
screw through the display and into the TRAM chute to
securely lock the TRAM chute to the display. Serious
injury or death could result if this warning is not
observed.
Follow these instructions to replace a TRAM chute and module with a
new TRAM chute and module.
1. Remove the TRAM module.
a. Disconnect the TRAM communication cable.
b. Press and hold the release levers simultaneously and pull the
module out about 15 cm (6 in).
c.
Once released, grasp the module firmly with both hands and
remove the rest of the way. Do not try to hold the module by the
release levers.
2. Remove the TRAM chute from the back of the display.
a. Tilt the monitor back and use a Phillips-head screwdriver to
remove the TRAM chute mounting screw from the foot.
b. Slide the TRAM chute to the right to release the mounting tabs
from the display.
3. Align the new, empty TRAM chute with the mounting slots.
Transport Pro
2012659-042
640A
8-19