Video Test Screens; Brightness Level Test; Speaker Test; Tram Communication Test - GE Transport Pro Service Manual

Patient monitor
Hide thumbs Also See for Transport Pro:
Table of Contents

Advertisement

Video Test Screens

Brightness Level Test

Speaker Test

TRAM Communication Test

Revision A
Checkout: Recommended Tests and Checkout Procedures
1. Access the Boot Loader program.
2. From the Service Menu, select Video Test Screens.
3. Select the first color test (White Screen), verify the display is correct
and press the TRIM KNOB control to exit the test.
4. Repeat for each of the tests:
Red Screen
Blue Screen
Green Screen
Vertical Bars
5. Select Exit.
6. Reboot the monitor.
1. Select MORE MENUS > MONITOR SETUP > BRIGHTNESS.
2. Verify the display brightness changes when setting the brightness
level from 100% to 10%. Reset the brightness level to your default
setting.
1. Select MORE MENUS > ALARM CONTROL > ALARM VOL.
2. Change the alarm volume to 100%.
3. Verify the speaker volume changes accordingly.
4. Return the volume to the level it was previously set to, before you
changed it for this test.
1. Confirm that all components of the monitoring system are correctly
connected as described in Chapter 3,
2. Connect the TRAM communication cable to the TRAM module and to
the monitor.
3. Turn on the monitor.
4. Verify that the Power indicator illuminates on the TRAM module.
5. Place the TRAM module into the chute.
6. Configure the display with as many waveforms as possible. Refer to
the operator's manual, if necessary.
7. Verify that the waveforms look clean (no noise).
Transport Pro
2012659-042
"Installation"
.
9-13

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents