Boot Loader Service Menu; Service Mode Menu; Review Errors - GE Transport Pro Service Manual

Patient monitor
Hide thumbs Also See for Transport Pro:
Table of Contents

Advertisement

Boot Loader Service Menu

Service Mode Menu

Review Errors

6-8
Troubleshooting: Service Menu
Use the Boot Loader service menu to download software. The Boot
Loader service menu contains the following options:
Menu Option
CHANGE SERVICE PORT
INTERNET ADDRESS
SET CONFIGURATION
VIDEO TEST SCREENS
BATTERY SIMULATION
WAKE UP BATTERY
OPTIONS MENU
Use the Service Mode menu to:
Configure the monitor.
Review errors.
Access battery information.
To access the Service Mode menu, select MORE MENUS > MONITOR
SETUP > SERVICE MODE and enter the password. The following
Service Mode menu options display.
Menu Option
REVIEW ERRORS
BATTERY SERVICE
PATIENT-MONITOR TYPE
MENU SETUP
MONITOR SETTINGS
The Review Errors menu is an advanced troubleshooting tool used by GE
engineering personnel to review error logs.
Transport Pro
2012659-042
Description
Change the monitor's IP address for software
upgrades.
Select country (not for text translation) and language
setting for displayed text.
Test display's color resolution.
For GE engineering use only.
Wake up a battery.
Clear the configuration memory (erases all patient
histories and reset all defaults to the factory installed
defaults).
Description
Review input and output error information (GE
engineering only).
Review battery data for the internal monitor batteries.
Select the type of monitor (i.e adult, neonatal or
operating room).
Define if a password is required to access the Monitor
Defaults menu.
Set the unit name and bed number.
Revision A