ePort Flex Circuit

Revision A
Disassembly: FRU Replacement
1. Remove the back plastic. See
2. Remove the two screws that connect the flex circuit cable to the main
processor PCB shield.
3. Remove the battery housing. See
Subassembly"
on page 8-9.
4. Remove the two screws attaching the flex circuit cable to the back
plastic.
5. Replace the flex circuit.
6. Assemble the back plastic. See
Transport Pro
2012659-042
"Back Plastic Disassembly"
"Main Processor PCB
"Back Plastic Assembly"
on page 8-4.