Inspection and Cleaning

Visual Inspection

Cleaning Cautions

5-4
Maintenance: Inspection and Cleaning
Remove power and all cables before inspecting or cleaning the
equipment.
Inspect the equipment and its components carefully prior to installation,
once every 12 months thereafter and each time the equipment is
serviced.
Carefully inspect the equipment for physical damage to the case, the
display and the controls. Do not use the equipment if damage is
determined. Refer damaged equipment to qualified service
personnel.
Inspect all external connections for loose connectors or frayed cables.
Have any damaged connectors or cables replaced by qualified service
personnel.
Inspect the display face for marks, scratches, or other damage.
Physical damage to a LCD touchscreen face may jeopardize its
performance. Have the LCD replaced by qualified service personnel if
necessary.
Safety labels and inscription on the device are clearly legible.
Check for loose or missing screws on the mounting hardware.
Improper cleaning methods can result in degradation of the equipment
performance and/or failure. To avoid damage to the equipment:
Never use conductive solutions, solutions that contain chlorides, wax,
or wax compounds to clean the equipment.
Never immerse equipment in any liquid.
Never pour any liquid on the equipment or permit fluid to seep into
connections or openings.
Never use the following solutions:
Acetone
Betadine
Methylene chloride
Methyl ethyl ketone
Hexane
Lysol
Sporicidin
Ammonia-based glass cleaners
Alcohol-based cleaning agents
Abrasive cleaners or solvents of any kind
Never autoclave or steam clean the equipment.
Transport Pro
2012659-042