Inverter Pcb - GE Transport Pro Service Manual

Patient monitor
Hide thumbs Also See for Transport Pro:
Table of Contents

Advertisement

Inverter PCB

Revision A
Disassembly: FRU Replacement
1. Remove the back plastic. See
2. Remove the display subassembly. See
page 8-7.
3. Remove the two screws holding the inverter board on the shield.
4. Remove the backlight friction connector from the main processor
PCB.
5. Install the new inverter board onto the shield.
6. Connect the backlight friction connector.
7. Secure the inverter board to the main processor PCB with the two
screws (PN 2000540-007).
8. Reinstall the display subassembly. Be sure to reconnect the backlight
friction, keypad PCB and display friction connectors to the main
processor PCB.
Transport Pro
2012659-042
"Back Plastic Assembly"
"Display Subassembly"
on page 8-6.
on
634A
635A
8-13