When About 5 Minutes Of Main Battery Charge Remains; When 5 Minutes Of Main Battery Charge Expires; After Plugging The Monitor Into Dc Power; E13 Battery Low - GE CareScape V100 Service Manual

Vital sings monitor
Hide thumbs Also See for CareScape V100:
Table of Contents

Advertisement

When about 5 minutes of main battery charge remains:

When 5 minutes of main battery charge expires:

After plugging the monitor into DC power:

E13 BATTERY LOW

3-4
Installation: Powering the monitor
The low-priority BATTERY LOW alarm escalates to a high-priority BATTERY LOW
alarm.
The BATTERY LOW indicator flashes.
Any NIBP determination in progress at the time of the alarm escalation is
allowed to finish.
Any Stat mode cycle that was initiated before the alarm escalation is
allowed to finish.
The user is not able to initiate:
any new NIBP determinations of any type
any printouts
NOTE
At this time, it is highly recommended to plug the monitor into external DC
power.
After 5 minutes of high-priority BATTERY LOW alarm, the monitor enters a
battery low shutdown.
No error code is displayed.
The BATTERY LOW indicator flashes.
The monitor alarms for 2.5 minutes, then shuts down completely.
NOTE
You must plug the monitor into DC power before resuming monitoring.
The BATTERY LOW indicator (when the monitor is on) and CHARGING
indicator illuminate.
The BATTERY LOW indicator turns off when the battery level reaches a
sufficient charge level to operate without the BATTERY LOW alarm active.
At any time while the high-priority BATTERY LOW alarm is active, certain actions
can trigger the 'E13' BATTERY LOW alarm: any attempt to start an NIBP
determination or a printout. This alarm is giving you additional warning that the
battery charge is critically low.
NOTE
At this time, it is highly recommended to plug the monitor into external DC
power.
The 'E13' error code appears in the min window.
The BATTERY LOW indicator flashes.
This alarm can be acknowledged by pressing the Silence button.
The user is not able to initiate:
any new NIBP determinations of any type
any printouts
CARESCAPE V100 Vital Signs Monitor