Introduction; General; Service Password; Boot Loader Access - GE Transport Pro Service Manual

Patient monitor
Hide thumbs Also See for Transport Pro:
Table of Contents

Advertisement

Introduction

General

Service Password

Boot Loader Access

Reboot

4-2
Configuration: Introduction
To configure the monitor, define each of the monitor settings listed in the
order given below.
Set Patient-Monitor Type.
Set Monitor Defaults Password.
Set Unit Name.
Set Bed Number.
Set Country Selection.
Set Language.
Refer to the operator's manual to configure the monitor for clinical use.
The service password is the date and month of the monitor (e.g. July 4 is
0407).
Many of the configuration settings are made in the Boot Loader program.
To access the Boot Loader program, follow this procedure.
1. Hold down NBP Go/Stop and Zero All.
2. Press and release the TRIM KNOB control.
3. Keep holding NBP Go/Stop and Zero All until the Boot Loader
program appears on the display.
Many of the configuration settings do not take effect until the monitor is
rebooted. To reboot the monitor, follow this procedure.
1. Hold down NBP Go/Stop and Zero All.
2. Press the TRIM KNOB control.
3. Release all when the display goes blank.
Transport Pro
2012659-042
Revision A

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Transport Pro and is the answer not in the manual?

Questions and answers

Table of Contents

Save PDF