Endress+Hauser Proline Promass E 300 Operating Instructions Manual
Endress+Hauser Proline Promass E 300 Operating Instructions Manual

Endress+Hauser Proline Promass E 300 Operating Instructions Manual

Coriolis flowmeter ethernet/ip
Hide thumbs Also See for Proline Promass E 300:

Advertisement

Quick Links

BA01727D/06/EN/03.24-00
71675064
2024-11-01
Valid as of version
01.00.zz (Device firmware)
Products
Operating Instructions
Proline Promass E 300
Coriolis flowmeter
EtherNet/IP
Solutions
Services

Advertisement

loading
Need help?

Need help?

Do you have a question about the Proline Promass E 300 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Endress+Hauser Proline Promass E 300

  • Page 1 Products Solutions Services BA01727D/06/EN/03.24-00 71675064 2024-11-01 Valid as of version 01.00.zz (Device firmware) Operating Instructions Proline Promass E 300 Coriolis flowmeter EtherNet/IP...
  • Page 2 • The manufacturer reserves the right to modify technical data without prior notice. Your Endress+Hauser sales organization will supply you with current information and updates to this manual. Endress+Hauser...
  • Page 3: Table Of Contents

    Proline Promass E 300 EtherNet/IP Table of contents Table of contents 6.1.2 Environmental and process About this document ....6 requirements ....
  • Page 4 Table of contents Proline Promass E 300 EtherNet/IP 8.3.11 Disabling write protection via access 10.6.5 Carrying out additional display code ......61 configurations .
  • Page 5 13.2 Measuring and test equipment ... 199 13.3 Endress+Hauser services ....199 Repair ......200 14.1 General notes .
  • Page 6: About This Document

    About this document Proline Promass E 300 EtherNet/IP About this document Document function These Operating Instructions contain all the information required in the various life cycle phases of the device: from product identification, incoming acceptance and storage, to installation, connection, operation and commissioning, through to troubleshooting, maintenance and disposal.
  • Page 7: Symbols For Certain Types Of Information

    Proline Promass E 300 EtherNet/IP About this document 1.2.4 Tool symbols Symbol Meaning Flat-blade screwdriver Allen key Open-ended wrench 1.2.5 Symbols for certain types of information Symbol Meaning Permitted Procedures, processes or actions that are permitted. Preferred Procedures, processes or actions that are preferred.
  • Page 8: Documentation

    • Device Viewer (www.endress.com/deviceviewer): Enter the serial number from the nameplate • Endress+Hauser Operations app: Enter serial number from nameplate or scan matrix code on nameplate. The following documentation may be available depending on the device version ordered: Document type...
  • Page 9: Safety Instructions

    Proline Promass E 300 EtherNet/IP Safety instructions Safety instructions Requirements for the personnel The personnel for installation, commissioning, diagnostics and maintenance must fulfill the following requirements: ‣ Trained, qualified specialists must have a relevant qualification for this specific function and task.
  • Page 10: Workplace Safety

    Verification for borderline cases: ‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability as minute changes in the temperature, concentration or level of contamination in the process can alter the corrosion resistance properties.
  • Page 11: Security

    Proline Promass E 300 EtherNet/IP Safety instructions IT security Our warranty is valid only if the product is installed and used as described in the Operating Instructions. The product is equipped with security mechanisms to protect it against any inadvertent changes to the settings.
  • Page 12: Infrastructure Mode

    Safety instructions Proline Promass E 300 EtherNet/IP User-specific access code Write access to the device parameters via the local display, web browser or operating tool (e.g. FieldCare, DeviceCare) can be protected by the modifiable, user-specific access code (→  145).
  • Page 13 Proline Promass E 300 EtherNet/IP Safety instructions This includes organizational security measures such as the assignment of access authorization as well as technical measures such as network segmentation. Transmitters with an Ex de approval may not be connected via the service interface (CDI-RJ45)! Order code for "Approval transmitter + sensor", options (Ex de): BA, BB, C1, C2, GA,...
  • Page 14: Product Description

    Product description Proline Promass E 300 EtherNet/IP Product description The device consists of a transmitter and a sensor. The device is available as a compact version: The transmitter and sensor form a mechanical unit. Product design N ic te r e rg ö...
  • Page 15: Incoming Acceptance And Product

    • Enter the serial numbers from the nameplates in the Device Viewer (www.endress.com/deviceviewer): all the information about the device is displayed. • Enter the serial numbers from the nameplates into the Endress+Hauser Operations app or scan the DataMatrix code on the nameplate with the Endress+Hauser Operations app: all the information about the device is displayed.
  • Page 16: Transmitter Nameplate

    Incoming acceptance and product identification Proline Promass E 300 EtherNet/IP 4.2.1 Transmitter nameplate 3 4 5 Order code: Ser. no.: Ext. ord. cd.: Date: A0029192  2 Example of a transmitter nameplate Manufacturer address/certificate holder Name of the transmitter Order code...
  • Page 17: Sensor Nameplate

    Proline Promass E 300 EtherNet/IP Incoming acceptance and product identification 4.2.2 Sensor nameplate Order code: Ser. no.: Ext. ord. cd.: A0029206  3 Example of a sensor nameplate, part 1 Name of the sensor Manufacturer address/certificate holder Nominal diameter of the sensor; flange nominal diameter/nominal pressure; sensor test pressure; medium temperature range;...
  • Page 18: Symbols On The Device

    Incoming acceptance and product identification Proline Promass E 300 EtherNet/IP 4.2.3 Symbols on the device Symbol Meaning WARNING! This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury. Please consult the documentation for the measuring instrument to discover the type of potential danger and measures to avoid it.
  • Page 19: Storage And Transport

    Proline Promass E 300 EtherNet/IP Storage and transport Storage and transport Storage conditions Observe the following notes for storage: ‣ Store in the original packaging to ensure protection from shock. ‣ Do not remove protective covers or protective caps installed on process connections.
  • Page 20: Measuring Devices With Lifting Lugs

    Storage and transport Proline Promass E 300 EtherNet/IP 5.2.2 Measuring devices with lifting lugs CAUTION Special transportation instructions for devices with lifting lugs ‣ Only use the lifting lugs fitted on the device or flanges to transport the device. ‣...
  • Page 21: Mounting

    Proline Promass E 300 EtherNet/IP Mounting Mounting Mounting requirements 6.1.1 Installation position Installation point A0028772 To prevent measuring errors arising from accumulation of gas bubbles in the measuring pipe, avoid the following mounting locations in the piping: • Highest point of a pipeline.
  • Page 22 Mounting Proline Promass E 300 EtherNet/IP Ø orifice plate, pipe restriction [mm] [in] [mm] [in] ³⁄₈ 0.24 ½ 0.40 0.55 1 ½ 0.87 1.10 1.97 Orientation The direction of the arrow on the sensor nameplate helps you to install the sensor according to the flow direction (direction of medium flow through the piping).
  • Page 23: Inlet And Outlet Runs

    Proline Promass E 300 EtherNet/IP Mounting Inlet and outlet runs No special precautions need to be taken for fittings that create turbulence, such as valves, elbows or T-pieces, as long as no cavitation occurs →  23. A0029322 A0029323 Installation dimensions For the dimensions and installation lengths of the device, see the "Technical...
  • Page 24 Mounting Proline Promass E 300 EtherNet/IP Thermal insulation In the case of some fluids, it is important to keep the heat radiated from the sensor to the transmitter to a low level. A wide range of materials can be used for the required insulation.
  • Page 25: Special Installation Instructions

    Proline Promass E 300 EtherNet/IP Mounting Vibrations The high oscillation frequency of the measuring tubes ensures that the correct operation of the measuring system is not influenced by plant vibrations. 6.1.3 Special installation instructions Drainability When installed vertically, the measuring tubes can be drained completely and protected against buildup.
  • Page 26: Weather Protection Cover

    Mounting Proline Promass E 300 EtherNet/IP Experience shows that zero adjustment is advisable only in special cases: • To achieve maximum measurement accuracy even with low flow rates. • Under extreme process or operating conditions (e.g. very high process temperatures or very high-viscosity fluids).
  • Page 27: Mounting The Measuring Instrument

    Proline Promass E 300 EtherNet/IP Mounting SW 2.5 5 (0.2) min. 15 (0.6) A0029800 Cover borehole for the securing screw Securing screw to lock the cover Mounting the measuring instrument 6.2.1 Required tools For sensor For flanges and other process connections: Use a suitable mounting tool.
  • Page 28: Turning The Transmitter Housing

    Mounting Proline Promass E 300 EtherNet/IP 2. Install the measuring device or turn the transmitter housing so that the cable entries do not point upwards.  A0029263 6.2.4 Turning the transmitter housing To provide easier access to the connection compartment or display module, the transmitter housing can be turned.
  • Page 29: Turning The Display Module

    Proline Promass E 300 EtherNet/IP Mounting 4 mm 7 Nm (5.2 lbf ) A0043150  10 Ex housing 1. Loosen the fixing screws. 2. Turn the housing to the desired position. 3. Tighten the securing screws. 6.2.5 Turning the display module The display module can be turned to optimize display readability and operability.
  • Page 30: Post-Installation Check

    Mounting Proline Promass E 300 EtherNet/IP Post-installation check Is the device undamaged (visual inspection)?  Does the measuring instrument correspond to the measuring point specifications? For example: • Process temperature →  221  • Pressure (refer to the "Pressure-temperature ratings" section of the "Technical Information"...
  • Page 31: Electrical Connection

    Proline Promass E 300 EtherNet/IP Electrical connection Electrical connection WARNING Live parts! Incorrect work performed on the electrical connections can result in an electric shock. ‣ Set up a disconnecting device (switch or power-circuit breaker) to easily disconnect the device from the supply voltage.
  • Page 32 Electrical connection Proline Promass E 300 EtherNet/IP Ethernet-APL Shielded twisted-pair cable. Cable type A is recommended. https://www.profibus.com Ethernet-APL White Paper " Current output 0 /4 to 20 mA (excluding HART) Standard installation cable is sufficient. Pulse /frequency /switch output Standard installation cable is sufficient.
  • Page 33: Endress+Hauser

    Proline Promass E 300 EtherNet/IP Electrical connection A standard cable with the following minimum requirements can be used as the connecting cable, even in the hazardous area (Zone 2, Class I, Division 2 and Zone 1, Class I, Division Standard cable 4 wires (2 pairs);...
  • Page 34: Terminal Assignment

    Electrical connection Proline Promass E 300 EtherNet/IP 7.2.3 Terminal assignment Transmitter: supply voltage, input/outputs The terminal assignment of the inputs and outputs depends on the individual order version of the device. The device-specific terminal assignment is documented on an adhesive label in the terminal cover.
  • Page 35: Connecting The Measuring Instrument

    Proline Promass E 300 EtherNet/IP Electrical connection 2. If the measuring device is supplied without cable glands: Provide suitable cable gland for corresponding connecting cable. 3. If the measuring device is supplied with cable glands: Observe requirements for connecting cables →  31.
  • Page 36 Electrical connection Proline Promass E 300 EtherNet/IP N ic N ic te r te r e rg ö ff e rg ö ff iz e iz e te n te n ri r ri r s io s io te r ö...
  • Page 37 Proline Promass E 300 EtherNet/IP Electrical connection 22 mm 24 mm – A0033984 4. Connect the cable according to the terminal assignment.  Signal cable terminal assignment: The device-specific terminal assignment is documented on an adhesive label in the terminal cover.
  • Page 38: Integrating The Transmitter Into A Network

    Electrical connection Proline Promass E 300 EtherNet/IP 7.3.2 Integrating the transmitter into a network This section only presents the basic options for integrating the device into a network. For information on the procedure to follow to connect the transmitter correctly →  35.
  • Page 39 Proline Promass E 300 EtherNet/IP Electrical connection Integrating into a ring topology The device is integrated via the terminal connection for signal transmission (output 1) and the connection to the service interface (CDI-RJ45). Note the following when connecting: • Recommended cable: CAT5e, CAT6 or CAT7, with shielded connector (e.g. brand: YAMAICHI ;...
  • Page 40: Connecting The Remote Display And Operating Module Dkx001

    Electrical connection Proline Promass E 300 EtherNet/IP 7.3.3 Connecting the remote display and operating module DKX001 The remote display and operating module DKX001 is available as an optional extra →  202.. • The remote display and operating module DKX001 is only available for the following housing version: order code for "Housing": option A "Aluminum, coated"...
  • Page 41: Special Connection Instructions

    Proline Promass E 300 EtherNet/IP Electrical connection Special connection instructions 7.5.1 Connection examples EtherNet/IP A0028767  12 Connection example for EtherNet/IP Control system (e.g. PLC) Ethernet switch Observe cable specifications Device plug Transmitter EtherNet/IP: DLR (Device Level Ring) A0027544 Control system (e.g. PLC) Ethernet switch Observe cable specifications →...
  • Page 42: Pulse/Frequency Output

    Electrical connection Proline Promass E 300 EtherNet/IP 4...20 mA A0028759  14 Connection example for 4-20 mA current output (passive) Automation system with current input (e.g. PLC) Active barrier for power supply (e.g. RN221N) Analog display unit: observe maximum load...
  • Page 43: Relay Output

    Proline Promass E 300 EtherNet/IP Electrical connection Relay output A0028760  17 Connection example for relay output (passive) Automation system with relay input (e.g. PLC) Power supply Transmitter: observe input values →  211 Current input A0028915  18 Connection example for 4 to 20 mA current input...
  • Page 44: Hardware Settings

    Electrical connection Proline Promass E 300 EtherNet/IP Hardware settings 7.6.1 Setting the device address The IP address of the measuring device can be configured for the network via DIP switches. Addressing data IP address and configuration options 1st octet 2nd octet...
  • Page 45: Activating The Default Ip Address

    Proline Promass E 300 EtherNet/IP Electrical connection 7.6.2 Activating the default IP address Activating the default IP address by DIP switch Risk of electric shock when opening the transmitter housing. ‣ Before opening the transmitter housing: ‣ Disconnect the device from the power supply.
  • Page 46: Post-Connection Check

    Electrical connection Proline Promass E 300 EtherNet/IP Post-connection check Are the device and cable undamaged (visual inspection)?  Is the protective earthing established correctly?  Do the cables used comply with the requirements ?  Are the installed cables strain-relieved and securely routed? ...
  • Page 47: Operation Options

    Proline Promass E 300 EtherNet/IP Operation options Operation options Overview of operation options A0034513 Local operation via display module Computer with web browser or operating tool (e.g. FieldCare, DeviceCare, AMS Device Manager, SIMATIC PDM) Field Xpert SFX350 or SFX370 Field Xpert SMT70 Mobile handheld terminal Automation system (e.g.
  • Page 48: Structure And Function Of The Operating Menu

    Operation options Proline Promass E 300 EtherNet/IP Structure and function of the operating menu 8.2.1 Structure of the operating menu For an overview of the operating menu for experts: see the "Description of Device Parameters" document supplied with the device...
  • Page 49: Operating Philosophy

    Proline Promass E 300 EtherNet/IP Operation options 8.2.2 Operating philosophy The individual parts of the operating menu are assigned to certain user roles (e.g. operator, maintenance etc.). Each user role contains typical tasks within the device life cycle. Menu/parameter User role and tasks...
  • Page 50: Access To Operating Menu Via Local Display

    Operation options Proline Promass E 300 EtherNet/IP Menu/parameter User role and tasks Content/meaning Expert Function- Tasks that require detailed knowledge of Contains all of the device parameters and allows direct access to these by oriented the function of the device: means of an access code.
  • Page 51: Display Area

    Proline Promass E 300 EtherNet/IP Operation options Display area In the display area, each measured value is prefaced by certain symbol types for further description: Measured variable Measurement channel Diagnostic behavior number ↓ ↓ ↓ Example Appears only if a diagnostics event is present for this measured variable.
  • Page 52 Operation options Proline Promass E 300 EtherNet/IP Diagnostic behavior Symbol Meaning Alarm • Measurement is interrupted. • Signal outputs and totalizers assume the defined alarm condition. • A diagnostic message is generated. Warning • Measurement is resumed. • The signal outputs and totalizers are not affected.
  • Page 53: Navigation View

    Proline Promass E 300 EtherNet/IP Operation options 8.3.2 Navigation view In the submenu In the wizard /../Operation 0091-1 /../Select medium Access stat.disp Select medium Operator Liquid Locking status Display A0013993-EN A0013995-EN Navigation view Navigation path to current position Status area Display area for navigation Operating elements →...
  • Page 54: Locking Procedure

    Operation options Proline Promass E 300 EtherNet/IP Setup Is displayed: • In the menu next to the "Setup" selection • At the left in the navigation path in the Setup menu Diagnosis Is displayed: • In the menu next to the "Diagnostics" selection •...
  • Page 55: Editing View

    Proline Promass E 300 EtherNet/IP Operation options 8.3.3 Editing view Numeric editor +0.000 Xx – A0034250  21 For entering values in parameters (e.g. limit values) Entry display area Input screen Confirm, delete or reject entry Operating elements Text editor...
  • Page 56 Operation options Proline Promass E 300 EtherNet/IP Operating key Meaning Enter key • Pressing the key briefly confirms your selection. • Pressing the key for 2 s confirms your entry. Escape key combination (press keys simultaneously) Close the editing view without accepting a change.
  • Page 57: Operating Elements

    Proline Promass E 300 EtherNet/IP Operation options 8.3.4 Operating elements Operating key Meaning Minus key In menu, submenu Moves the selection bar upwards in a picklist In wizards Goes to previous parameter In the text and numeric editor Move the entry position to the left.
  • Page 58 Operation options Proline Promass E 300 EtherNet/IP Calling up and closing the context menu The user is in the operational display. 1. Press the  and  keys for longer than 3 seconds.  The context menu opens. XXXXXXXXXX Setup Conf.backup...
  • Page 59: Navigating And Selecting From List

    Proline Promass E 300 EtherNet/IP Operation options 8.3.6 Navigating and selecting from list Different operating elements are used to navigate through the operating menu. The navigation path is displayed on the left in the header. Icons are displayed in front of the individual menus.
  • Page 60: Calling Up Help Text

    Operation options Proline Promass E 300 EtherNet/IP The direct access code consists of a 5-digit number (at maximum) and the channel number, which identifies the channel of a process variable: e.g. 00914-2. In the navigation view, this appears on the right-hand side in the header of the selected parameter.
  • Page 61: User Roles And Related Access Authorization

    Proline Promass E 300 EtherNet/IP Operation options Ent. access code Invalid or out of range input value Min:0 Max:9999 A0014049-EN For a description of the editing view - consisting of the text editor and numeric editor - with symbols →  55, for a description of the operating elements →  57 8.3.10...
  • Page 62: Enabling And Disabling The Keypad

    Operation options Proline Promass E 300 EtherNet/IP 2. Enter the access code.  The -symbol in front of the parameters disappears; all previously write- protected parameters are now re-enabled. 8.3.12 Enabling and disabling the keypad lock The keypad lock makes it possible to block access to the entire operating menu via local operation.
  • Page 63: Requirements

    Proline Promass E 300 EtherNet/IP Operation options 8.4.2 Requirements Computer hardware Hardware Interface CDI-RJ45 WLAN Interface The computer must have a RJ45 The operating unit must have a WLAN interface. interface. Connection Standard Ethernet cable Connection via Wireless LAN. Screen Recommended size: ≥12"...
  • Page 64: Connecting The Device

    Operation options Proline Promass E 300 EtherNet/IP Measuring device: Via CDI-RJ45 service interface Device CDI-RJ45 service interface Measuring device The measuring device has an RJ45 interface. Web server Web server must be enabled; factory setting: ON  For information on enabling the Web server →  68...
  • Page 65 Proline Promass E 300 EtherNet/IP Operation options Configuring the Internet protocol of the computer The IP address can be assigned to the measuring device in a variety of ways: • Dynamic Host Configuration Protocol (DHCP), factory setting: The IP address is automatically assigned to the measuring device by the automation system (DHCP server).
  • Page 66: Enter The Password

    Operation options Proline Promass E 300 EtherNet/IP Establishing a WLAN connection from the mobile terminal to the measuring device 1. In the WLAN settings of the mobile terminal: Select the measuring device using the SSID (e.g. EH_Promass_300_A802000). 2. If necessary, select the WPA2 encryption method.
  • Page 67: Logging On

    Proline Promass E 300 EtherNet/IP Operation options 8.4.4 Logging on 1. Select the preferred operating language for the Web browser. 2. Enter the user-specific access code. 3. Press OK to confirm your entry. Access code 0000 (factory setting); can be changed by customer If no action is performed for 10 minutes, the Web browser automatically returns to the login page.
  • Page 68: Disabling The Web Server

    Operation options Proline Promass E 300 EtherNet/IP Functions Meaning Data exchange between computer and measuring device: • Device configuration: • Load settings from the device (XML format, save configuration) • Save settings to the device (XML format, restore configuration) • Logbook - Export Event logbook (.csv file) Data •...
  • Page 69: Logging Out

    Proline Promass E 300 EtherNet/IP Operation options Function scope of the "Web server functionality" parameter Option Description • The Web server is completely disabled. • Port 80 is locked. HTML Off The HTML version of the Web server is not available.
  • Page 70: Star Topology

    Operation options Proline Promass E 300 EtherNet/IP Star topology A0032078  24 Options for remote operation via EtherNet/IP network: star topology Automation system, e.g. "RSLogix" (Rockwell Automation) Workstation for measuring device operation: with Custom Add-On Profile for "RSLogix 5000" (Rockwell Automation) or with Electronic Data Sheet (EDS) Computer with Web browser (e.g.
  • Page 71: Service Interface

    Proline Promass E 300 EtherNet/IP Operation options Service interface Via service interface (CDI-RJ45) A point-to-point connection can be established to configure the device onsite. With the housing open, the connection is established directly via the service interface (CDI-RJ45) of the device.
  • Page 72 Operation options Proline Promass E 300 EtherNet/IP A0034570 Transmitter with integrated WLAN antenna Transmitter with external WLAN antenna LED lit constantly: WLAN reception is enabled on measuring device LED flashing: WLAN connection established between operating unit and measuring device Computer with WLAN interface and web browser (e.g. Microsoft Internet Explorer, Microsoft Edge) for accessing the integrated device web server or with operating tool (e.g.
  • Page 73: Fieldcare

    FieldCare Function range FDT-based (Field Device Technology) plant asset management tool from Endress+Hauser. It can configure all smart field units in a system and helps you manage them. By using the status information, it is also a simple but effective way of checking their status and condition.
  • Page 74: Devicecare

    DeviceCare Function range Tool for connecting and configuring Endress+Hauser field devices. The fastest way to configure Endress+Hauser field devices is with the dedicated "DeviceCare" tool. Together with the device type managers (DTMs) it presents a convenient, comprehensive solution. Innovation brochure IN01047S Source for device description files →...
  • Page 75: System Integration

    Proline Promass E 300 EtherNet/IP System integration System integration Overview of device description files 9.1.1 Current version data for the device Firmware version 01.00.zz • On the title page of the manual • On the transmitter nameplate • Firmware version Diagnostics →...
  • Page 76: Integrating The Measuring Device In The

    System integration Proline Promass E 300 EtherNet/IP Integrating the measuring device in the system For detailed information on system integration, see the Operating Instructions for the device A detailed description of how to integrate the device into an automation system (e.g.
  • Page 77: Input And Output Groups

    Proline Promass E 300 EtherNet/IP System integration 9.4.2 Input and output groups Possible configurations Configuration 1: Exclusive Owner Multicast Input Assembly Fix Instance Size [byte] Min. RPI (ms) Input Assembly Configurable Configuration 0 x 64 – Output Assembly Fix O → T Configuration...
  • Page 78: Possible Connections

    System integration Proline Promass E 300 EtherNet/IP Configuration 7: Exclusive Owner Multicast Input Assembly Configurable Instance Size [byte] Min. RPI (ms) Input Assembly Configurable Configuration 0 x 69 – – Output Assembly Fix O → T Configuration 0 x 66 Input Assembly Fix T →...
  • Page 79 Proline Promass E 300 EtherNet/IP System integration Description Byte Totalizer 2 37 to 40 Totalizer 3 41 to 44 Diagnostic information via EtherNet/IP →  87 Mass flow fixed input assembly (Assem106), 32 byte Description Byte 1. File header (not visible) 1 to 4 2.
  • Page 80 System integration Proline Promass E 300 EtherNet/IP Description Byte 8. Concentration 57 to 60 9. Volume flow unit 61 to 62 10. Corrected volume flow unit 63 to 64 11. Concentration unit 65 to 66 Only available with the Concentration application package.
  • Page 81 Proline Promass E 300 EtherNet/IP System integration Description Byte 3. Oscillation frequency 0 37 to 40 4. Oscillation frequency 1 41 to 44 5. Oscillation amplitude 0 45 to 48 6. Oscillation amplitude 1 49 to 52 7. Oscillation damping 0 53 to 56 8.
  • Page 82 System integration Proline Promass E 300 EtherNet/IP Possible input values 11 to 20: • Off • Temperature unit • Totalizer 1 unit • Current diagnostics • Density unit • Totalizer 2 unit • Previous diagnostics • Reference density unit • Totalizer 3 unit •...
  • Page 83 Proline Promass E 300 EtherNet/IP System integration Description (format) Byte Value 22. Select concentration of medium type 13 to • 3062 (0) : Aqueous Fructose • 3063 (0) : Aqueous Glucose • 3068 (0) : Aqueous Hydrochloric Acid • 3077 (0) : Aqueous Hydrogen Peroxide •...
  • Page 84 System integration Proline Promass E 300 EtherNet/IP Description (format) Byte Value 32. Not Used 39 to – 33. External value % S&W (real) 41 to Data format: Byte 1-4: External value, % S&W Floating-point number (IEEE754) 34. External value, water cut % (real)
  • Page 85 Proline Promass E 300 EtherNet/IP System integration Description (format) Bits Byte Offset Parameter 244 Totalizer operation Preset value tot. 1 Parameter 97 Totalizer operation Control totalizer 1 Parameter 95 Totalizer 2 Assign process variable Parameter 107 Totalizer 2 Unit totalizer...
  • Page 86 System integration Proline Promass E 300 EtherNet/IP Description (format) Bits Byte Offset Parameter 189 Partially filled pipe detection High value partial filled pipe detection Parameter 190 Empty pipe detection Response time part. filled pipe detect. Parameter 182 Corrected volume flow calculation...
  • Page 87: Diagnostic Information Via Ethernet/Ip

    Proline Promass E 300 EtherNet/IP System integration Description (format) Bits Byte Offset 104. Parameter 67 Diagnostic behavior Assign behavior for diagnostic information 835 105. Parameter 72 Diagnostic behavior Assign behavior for diagnostic information 862 106. Parameter 68 Diagnostic behavior Assign behavior for diagnostic information 912 107.
  • Page 88 System integration Proline Promass E 300 EtherNet/IP Status signal Short text Value Tubes not oscillating 16777296 Configuration incompatible 16777312 Software incompatible 16777319 Modules incompatible 16777323 Main electronic failure 16777337 Main electronic failure 16777340 Main electronic failure 16777341 Main electronic failure...
  • Page 89 Proline Promass E 300 EtherNet/IP System integration Status signal Short text Value Device failure 16777547 Special event 10 16777552 Special event 11 16777553 Special event 12 16777554 Data storage 16777581 Memory content 16777582 Memory content 16777583 Measuring error too high...
  • Page 90 System integration Proline Promass E 300 EtherNet/IP Status signal Short text Value Process temperature too high 134217925 Process temperature too low 134217926 Sensor limit exceeded 134217928 Sensor limit exceeded 134217930 Sensor signal 134217932 Medium unsuitable 134217933 Main electronic failure 134217934...
  • Page 91 Proline Promass E 300 EtherNet/IP System integration Status signal Short text Value 1446 Device verification active 268435902 1447 Record application reference data 268435903 1448 Application reference data recorded 268435904 1449 Recording application ref. data failed 268435905 1450 Monitoring off 268435906...
  • Page 92: Commissioning

    Commissioning Proline Promass E 300 EtherNet/IP Commissioning 10.1 Post-mounting and post-connection check Before commissioning the device: ‣ Make sure that the post-installation and post-connection checks have been performed successfully. • Checklist for "Post-installation" check→  30 • Checklist for "Post-connection" check→  46 10.2...
  • Page 93: Configuring The Measuring Instrument

    Proline Promass E 300 EtherNet/IP Commissioning 10.5 Configuring the measuring instrument The Setup menu with its guided wizards contains all the parameters needed for standard operation. X X X X X X X 20.50 Main menu Display language English Display/operat.
  • Page 94: Defining The Tag Name

    Commissioning Proline Promass E 300 EtherNet/IP ‣ Display →  116 ‣ Low flow cut off →  120 ‣ →  121 Partially filled pipe detection ‣ Advanced setup →  122 10.5.1 Defining the tag name To enable fast identification of the measuring point within the system, you can enter a unique designation using the Device tag parameter and thus change the factory setting.
  • Page 95: Parameter Overview With Brief Description

    Proline Promass E 300 EtherNet/IP Commissioning Volume flow unit →  95 Volume unit →  95 →  95 Corrected volume flow unit Corrected volume unit →  95 Density unit →  95 Reference density unit →  96 →...
  • Page 96: Configuring The Communication Interface

    Commissioning Proline Promass E 300 EtherNet/IP Parameter Description Selection Factory setting Reference density unit Select reference density unit. Unit choose list Country-specific • kg/Nl • lb/Sft³ Density 2 unit Select second density unit. Unit choose list Country-specific: • kg/l • lb/ft³...
  • Page 97 Proline Promass E 300 EtherNet/IP Commissioning Parameter overview with brief description Parameter Description User interface / Selection /  Factory setting User entry MAC address Displays the MAC address of the measuring Unique 12-digit character Each measuring device is given device. string comprising letters and an individual address.
  • Page 98: Selecting And Setting The Medium

    Commissioning Proline Promass E 300 EtherNet/IP 10.5.4 Selecting and setting the medium The Select medium wizard submenu contains parameters that must be configured in order to select and set the medium. Navigation "Setup" menu → Medium selection ‣ Medium selection Select medium →...
  • Page 99: Displaying The I/O Configuration

    Proline Promass E 300 EtherNet/IP Commissioning Parameter Prerequisite Description Selection / User entry /  User interface Reference sound velocity In the Select gas type parameter, the Enter sound velocity of gas at 0 °C (32 1 to 99 999.9999 m/s Others option is selected.
  • Page 100: Configuring The Current Input

    Commissioning Proline Promass E 300 EtherNet/IP Parameter overview with brief description Parameter Description User interface / Selection / User entry I/O module 1 to n terminal numbers Shows the terminal numbers used by the I/O module. • Not used • 26-27 (I/O 1) • 24-25 (I/O 2) •...
  • Page 101: Configuring The Status Input

    Proline Promass E 300 EtherNet/IP Commissioning Parameter overview with brief description Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Terminal number – Shows the terminal numbers • Not used – used by the current input • 24-25 (I/O 2) module.
  • Page 102: Configuring The Current Output

    Commissioning Proline Promass E 300 EtherNet/IP Parameter overview with brief description Parameter Description Selection / User interface / User entry Assign status input Select function for the status input. • Off • Reset totalizer 1 • Reset totalizer 2 • Reset totalizer 3 • Reset all totalizers •...
  • Page 103 Proline Promass E 300 EtherNet/IP Commissioning Parameter overview with brief description Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Terminal number – Shows the terminal numbers • Not used – used by the current output • 24-25 (I/O 2) module.
  • Page 104 Commissioning Proline Promass E 300 EtherNet/IP Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Current span – Select current range for • 4...20 mA NAMUR Depends on country: process value output and • 4...20 mA US • 4...20 mA NAMUR upper/lower level for alarm •...
  • Page 105: Configuring The Pulse/Frequency/Switch Output

    Proline Promass E 300 EtherNet/IP Commissioning 10.5.9 Configuring the pulse/frequency/switch output The Pulse/frequency/switch output wizard guides you systematically through all the parameters that can be set for configuring the selected output type. Navigation "Setup" menu → Advanced setup → Pulse/frequency/switch output ‣...
  • Page 106 Commissioning Proline Promass E 300 EtherNet/IP Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting interface / User entry Operating mode – Define the output as a pulse, • Pulse – frequency or switch output. • Frequency • Switch Terminal number –...
  • Page 107 Proline Promass E 300 EtherNet/IP Commissioning Parameter Prerequisite Description Selection / User Factory setting interface / User entry Failure mode The Pulse option is selected in Define output behavior in • Actual value – the Operating mode alarm condition. • No pulses parameter (→  105) and a...
  • Page 108 Commissioning Proline Promass E 300 EtherNet/IP Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting interface / User entry Operating mode – Define the output as a pulse, • Pulse – frequency or switch output. • Frequency • Switch Terminal number –...
  • Page 109 Proline Promass E 300 EtherNet/IP Commissioning Parameter Prerequisite Description Selection / User Factory setting interface / User entry Assign frequency output The Frequency option is Select process variable for • Off – selected in Operating mode frequency output. • Mass flow parameter (→  105).
  • Page 110 Commissioning Proline Promass E 300 EtherNet/IP Parameter Prerequisite Description Selection / User Factory setting interface / User entry Maximum frequency value The Frequency option is Enter maximum frequency. 0.0 to 10 000.0 Hz – selected in the Operating mode parameter (→  105) and a process variable is...
  • Page 111 Proline Promass E 300 EtherNet/IP Commissioning Configuring the switch output Navigation "Setup" menu → Pulse/frequency/switch output ‣ Pulse/frequency/switch output 1 to n Operating mode →  111 Terminal number →  111 Signal mode →  111 Switch output function →...
  • Page 112 Commissioning Proline Promass E 300 EtherNet/IP Parameter Prerequisite Description Selection / User Factory setting interface / User entry Switch output function The Switch option is selected Select function for switch • Off – in the Operating mode output. • On parameter. • Diagnostic behavior •...
  • Page 113: 10.5.10 Configuring The Relay Output

    Proline Promass E 300 EtherNet/IP Commissioning Parameter Prerequisite Description Selection / User Factory setting interface / User entry Assign flow direction check • The Switch option is Select process variable for flow – selected in the Operating direction monitoring. mode parameter. • The Flow direction check...
  • Page 114 Commissioning Proline Promass E 300 EtherNet/IP Relay output function →  114 Assign flow direction check →  114 →  115 Assign limit Assign diagnostic behavior →  115 Assign status →  115 Switch-off value →  115 →...
  • Page 115 Proline Promass E 300 EtherNet/IP Commissioning Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Assign limit The Limit option is selected in Select process variable for limit • Mass flow – Relay output function function. • Volume flow parameter.
  • Page 116: 10.5.11 Configuring The Local Display

    Commissioning Proline Promass E 300 EtherNet/IP Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Failure mode – Define output behavior in • Actual status – alarm condition. • Open • Closed Switch status – Shows the current relay switch •...
  • Page 117 Proline Promass E 300 EtherNet/IP Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Format display A local display is provided. Select how measured values • 1 value, max. size – are shown on the display.
  • Page 118 Commissioning Proline Promass E 300 EtherNet/IP Parameter Prerequisite Description Selection / User Factory setting entry Value 1 display A local display is provided. Select the measured value that • Mass flow – is shown on the local display. • Volume flow • Corrected volume flow •...
  • Page 119 Proline Promass E 300 EtherNet/IP Commissioning Parameter Prerequisite Description Selection / User Factory setting entry • Current output 2 • Current output 3 0% bargraph value 1 A local display is provided. Enter 0% value for bar graph Signed floating-point Country-specific: display.
  • Page 120: 10.5.12 Configuring The Low Flow Cut Off

    Commissioning Proline Promass E 300 EtherNet/IP 10.5.12 Configuring the low flow cut off The Low flow cut off wizard systematically guides the user through all the parameters that must be set to configure low flow cut off. Navigation "Setup" menu → Low flow cut off ‣...
  • Page 121: 10.5.13 Configuring Partially Filled Pipe Detection

    Proline Promass E 300 EtherNet/IP Commissioning 10.5.13 Configuring partially filled pipe detection The Partial filled pipe detection wizard guides you systematically through all parameters that have to be set for configuring the monitoring of the pipe filling. Navigation "Setup" menu → Partially filled pipe detection ‣...
  • Page 122: Advanced Settings

    Commissioning Proline Promass E 300 EtherNet/IP 10.6 Advanced settings The Advanced setup submenu with its submenus contains parameters for specific settings. Navigation to the "Advanced setup" submenu X X X X X X X 20.50 0104-1 Main menu Display language English Display/operat.
  • Page 123: Using The Parameter To Enter The Access Code

    Proline Promass E 300 EtherNet/IP Commissioning ‣ Totalizer 1 to n →  131 ‣ Display →  133 ‣ WLAN settings ‣ Concentration ‣ Heartbeat setup ‣ Configuration backup →  139 ‣ →  140 Administration 10.6.1 Using the parameter to enter the access code Navigation "Setup"...
  • Page 124 Commissioning Proline Promass E 300 EtherNet/IP "Corrected volume flow calculation" submenu Navigation "Setup" menu → Advanced setup → Calculated values → Corrected volume flow calculation ‣ Corrected volume flow calculation →  124 Corrected volume flow calculation (1812) External reference density (6198) →...
  • Page 125: Carrying Out A Sensor Adjustment

    Proline Promass E 300 EtherNet/IP Commissioning Parameter Prerequisite Description Selection / User Factory setting interface / User entry Linear expansion coefficient The Calculated reference Enter linear, medium-specific Signed floating-point – density option is selected in expansion coefficient for number the Corrected volume flow calculating the reference calculation parameter density.
  • Page 126 Commissioning Proline Promass E 300 EtherNet/IP Performing density adjustment Note the following before performing the adjustment: • A density adjustment only makes sense if there is little variation in the operating conditions and the density adjustment is performed under the operating conditions.
  • Page 127 Proline Promass E 300 EtherNet/IP Commissioning 4. Select the Measure density 1 option and confirm.  In the Execute density adjustment parameter the following options are now available: Measure density 2 Restore original 5. Select the Measure density 2 option and confirm.
  • Page 128 Commissioning Proline Promass E 300 EtherNet/IP Parameter Prerequisite Description Selection / User Factory setting entry / User interface Density setpoint 2 In the Density adjustment The entry depends on – mode parameter, the 2 point the unit selected in adjustment option is selected. the Density unit parameter (0555).
  • Page 129 Proline Promass E 300 EtherNet/IP Commissioning Navigation "Setup" menu → Advanced setup → Sensor adjustment → Zero verification ‣ Zero verification Process conditions →  129 Progress →  129 →  129 Status Additional information →  129 Recommendation: →...
  • Page 130 Commissioning Proline Promass E 300 EtherNet/IP Parameter Description Selection / User interface Factory setting Zero point measured Shows the zero point measured for the Signed floating-point number – adjustment. Zero point standard deviation Shows the standard deviation of the zero Positive floating-point number –...
  • Page 131: Configuring The Totalizer

    Proline Promass E 300 EtherNet/IP Commissioning Parameter overview with brief description Parameter Description Selection / User interface Factory setting Process conditions Ensure process conditions as follows. • Tubes are completely filled – • Process operational pressure applied • No-flow conditions (closed valves) •...
  • Page 132 Commissioning Proline Promass E 300 EtherNet/IP Parameter overview with brief description Parameter Prerequisite Description Selection Factory setting Assign process variable – Select process variable for • Off – totalizer. • Mass flow • Volume flow • Corrected volume flow • Target mass flow •...
  • Page 133: Carrying Out Additional Display Configurations

    Proline Promass E 300 EtherNet/IP Commissioning 10.6.5 Carrying out additional display configurations In the Display submenu you can set all the parameters associated with the configuration of the local display. Navigation "Setup" menu → Advanced setup → Display ‣ Display Format display →...
  • Page 134 Commissioning Proline Promass E 300 EtherNet/IP Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Format display A local display is provided. Select how measured values • 1 value, max. size – are shown on the display.
  • Page 135 Proline Promass E 300 EtherNet/IP Commissioning Parameter Prerequisite Description Selection / User Factory setting entry Value 1 display A local display is provided. Select the measured value that • Mass flow – is shown on the local display. • Volume flow • Corrected volume flow •...
  • Page 136 Commissioning Proline Promass E 300 EtherNet/IP Parameter Prerequisite Description Selection / User Factory setting entry • Current output 2 • Current output 3 0% bargraph value 1 A local display is provided. Enter 0% value for bar graph Signed floating-point Country-specific: display.
  • Page 137: Wlan Configuration

    Proline Promass E 300 EtherNet/IP Commissioning Parameter Prerequisite Description Selection / User Factory setting entry Display language A local display is provided. Set display language. • English English • Deutsch (alternatively, the • Français ordered language is • Español preset in the device) •...
  • Page 138 Commissioning Proline Promass E 300 EtherNet/IP Navigation "Setup" menu → Advanced setup → WLAN settings ‣ WLAN settings WLAN →  138 WLAN mode →  138 →  138 SSID name Network security →  138 Security identification →  139 User name →...
  • Page 139: Configuration Management

    Proline Promass E 300 EtherNet/IP Commissioning Parameter Prerequisite Description Selection / User Factory setting entry / User interface Security identification – Select security settings and • Root certificate – download these settings via • Device certificate menu Data management > • Device private key Security >...
  • Page 140: Function Scope Of The "Configuration Management" Parameter

    Commissioning Proline Promass E 300 EtherNet/IP Backup state →  140 Comparison result →  140 Parameter overview with brief description Parameter Description User interface / Selection Operating time Indicates how long the device has been in operation. Days (d), hours (h), minutes (m) and seconds...
  • Page 141 Proline Promass E 300 EtherNet/IP Commissioning Navigation "Setup" menu → Advanced setup → Administration ‣ Administration ‣ Define access code →  141 ‣ Reset access code →  141 →  142 Device reset Using the parameter to define the access code Navigation "Setup"...
  • Page 142: Simulation

    Reset access code to factory settings. Character string comprising numbers, letters and special characters  For a reset code, contact your Endress+Hauser service organization. The reset code can only be entered via: • Web browser • DeviceCare, FieldCare (via CDI-RJ45 service interface) •...
  • Page 143 Proline Promass E 300 EtherNet/IP Commissioning Value current output 1 to n →  144 →  144 Frequency output simulation 1 to n Frequency value 1 to n →  144 Pulse output simulation 1 to n →  144 Pulse value 1 to n →...
  • Page 144 Commissioning Proline Promass E 300 EtherNet/IP Parameter overview with brief description Parameter Prerequisite Description Selection / User entry Assign simulation process variable – Select a process variable for the • Off simulation process that is activated. • Mass flow • Volume flow •...
  • Page 145: Protecting Settings From Unauthorized Access

    Proline Promass E 300 EtherNet/IP Commissioning Parameter Prerequisite Description Selection / User entry Switch output simulation 1 to n In the Operating mode parameter, the Switch the simulation of the switch • Off Switch option is selected. output on and off. • On Switch status 1 to n –...
  • Page 146 Commissioning Proline Promass E 300 EtherNet/IP 3. Enter the access code again in the Confirm access code parameter (→  141) to confirm.  The  symbol appears in front of all write-protected parameters. • Disabling parameter write protection via access code →  61.
  • Page 147: Write Protection Via Write Protection Switch

    Commissioning Via Web browser, FieldCare, DeviceCare (via CDI-RJ45 service interface), fieldbus You can only obtain a reset code from your local Endress+Hauser service organization. The code must be calculated explicitly for every device. 1. Note down the serial number of the device.
  • Page 148 Commissioning Proline Promass E 300 EtherNet/IP 2. Setting the write protection (WP) switch on the main electronics module to the OFF position (factory setting) disables hardware write protection.  No option is displayed in the Locking status parameter →  149. On the local display, the ...
  • Page 149: Operation

    Proline Promass E 300 EtherNet/IP Operation Operation 11.1 Reading off the device locking status Device active write protection: Locking status parameter Operation → Locking status Function scope of the "Locking status" parameter Options Description None The access authorization displayed in the Access status parameter applies →...
  • Page 150: Measured Variables" Submenu

    Operation Proline Promass E 300 EtherNet/IP 11.4.1 "Measured variables" submenu The Measured variables submenu contains all the parameters needed to display the current measured values for each process variable. Navigation "Diagnostics" menu → Measured values → Measured variables ‣ Measured variables Mass flow →...
  • Page 151 Proline Promass E 300 EtherNet/IP Operation GSV flow →  155 →  155 GSV flow alternative NSV flow →  156 NSV flow alternative →  156 Oil CTL →  156 Oil CPL →  156 Oil CTPL →...
  • Page 152: Density

    Operation Proline Promass E 300 EtherNet/IP Parameter overview with brief description Parameter Prerequisite Description User interface Factory setting Mass flow – Displays the mass flow that is Signed floating-point – currently measured. number Dependency The unit is taken from: Mass flow unit parameter (→...
  • Page 153: Target Corrected Volume Flow

    Proline Promass E 300 EtherNet/IP Operation Parameter Prerequisite Description User interface Factory setting Target mass flow With the following conditions: Displays the mass flow that is Signed floating-point – Order code for "Application currently measured for the number package", option ED target medium.
  • Page 154: Ctl

    Operation Proline Promass E 300 EtherNet/IP Parameter Prerequisite Description User interface Factory setting Carrier volume flow With the following conditions: Displays the volume flow Signed floating-point – • Order code for "Application currently measured for the number package", option ED carrier medium.
  • Page 155 Proline Promass E 300 EtherNet/IP Operation Parameter Prerequisite Description User interface Factory setting S&W volume flow For the following order code: Displays the S&W volume flow Signed floating-point – • "Application package", option which is calculated from the number EJ "Petroleum"...
  • Page 156 Operation Proline Promass E 300 EtherNet/IP Parameter Prerequisite Description User interface Factory setting NSV flow For the following order code: Displays the net volume flow Signed floating-point – • "Application package", option which is calculated from the number EJ "Petroleum"...
  • Page 157 Proline Promass E 300 EtherNet/IP Operation Parameter Prerequisite Description User interface Factory setting Water CTL For the following order code: Displays the correction factor Positive floating- – • "Application package", option which represents the effect of point number EJ "Petroleum"...
  • Page 158 Operation Proline Promass E 300 EtherNet/IP Parameter Prerequisite Description User interface Factory setting Water reference density unit For the following order code: Signed floating-point – • "Application package", option number EJ "Petroleum" • In the Petroleum mode parameter, the Net oil &...
  • Page 159 Proline Promass E 300 EtherNet/IP Operation Parameter Prerequisite Description User interface Factory setting Oil corrected volume flow For the following order code: Displays the currently Signed floating-point – • "Application package", option calculated volume flow of the number EJ "Petroleum"...
  • Page 160: Totalizer" Submenu

    Operation Proline Promass E 300 EtherNet/IP Parameter Prerequisite Description User interface Factory setting Weighted density average For the following order code: Displays the weighted average Signed floating-point – • "Application package", option for the density since the last number EJ "Petroleum"...
  • Page 161 Proline Promass E 300 EtherNet/IP Operation Navigation "Diagnostics" menu → Measured values → Input values ‣ Input values ‣ Current input 1 to n →  161 ‣ Status input 1 to n →  161 Input values of current input The Current input 1 to n submenu contains all the parameters needed to display the current measured values for every current input.
  • Page 162: Output Values

    Operation Proline Promass E 300 EtherNet/IP 11.4.4 Output values The Output values submenu contains all the parameters needed to display the current measured values for every output. Navigation "Diagnostics" menu → Measured values → Output values ‣ Output values ‣...
  • Page 163 Proline Promass E 300 EtherNet/IP Operation Navigation "Diagnostics" menu → Measured values → Output values → Pulse/frequency/switch output 1 to n ‣ Pulse/frequency/switch output 1 to n →  163 Output frequency 1 to n Pulse output 1 to n →...
  • Page 164: Adapting The Measuring Device To The Process Conditions

    Operation Proline Promass E 300 EtherNet/IP 11.5 Adapting the measuring device to the process conditions The following are available for this purpose: • Basic settings using theSetup menu (→  93) • Advanced settings using theAdvanced setup submenu (→  122) 11.6...
  • Page 165: Function Scope Of "Control Totalizer" Parameter

    Proline Promass E 300 EtherNet/IP Operation Parameter Prerequisite Description Selection / User Factory setting entry / User interface Weighted density average For the following order code: Displays the weighted average Signed floating-point – • "Application package", option for the density since the last number EJ "Petroleum"...
  • Page 166: Function Range Of "Reset All Totalizers" Parameter

    Operation Proline Promass E 300 EtherNet/IP 11.6.2 Function range of "Reset all totalizers" parameter Options Description Cancel No action is executed and the user exits the parameter. Reset + totalize Resets all totalizers to 0 and restarts the totaling process. This deletes all the previously aggregated flow values.
  • Page 167 Proline Promass E 300 EtherNet/IP Operation Clear logging data →  169 →  169 Data logging Logging delay →  169 Data logging control →  169 Data logging status →  169 Entire logging duration →  169 ‣...
  • Page 168 Operation Proline Promass E 300 EtherNet/IP Parameter overview with brief description Parameter Prerequisite Description Selection / User entry /  User interface Assign channel 1 The Extended HistoROM application Assign process variable to logging • Off package is available. channel. • Mass flow • Volume flow •...
  • Page 169 Proline Promass E 300 EtherNet/IP Operation Parameter Prerequisite Description Selection / User entry /  User interface Assign channel 3 The Extended HistoROM application Assign a process variable to logging For the picklist, see Assign package is available. channel. channel 1 parameter (→  168) ...
  • Page 170: Diagnostics And Troubleshooting

    Diagnostics and troubleshooting Proline Promass E 300 EtherNet/IP Diagnostics and troubleshooting 12.1 General troubleshooting For local display Error Possible causes Remedial action Local display is dark, but signal output is within The cable of the display module is not plugged in...
  • Page 171 Proline Promass E 300 EtherNet/IP Diagnostics and troubleshooting For access Fault Possible causes Remedial action Write access to parameters is not possible. Hardware write protection is enabled. Set the write protection switch on the main electronics module to the OFF position →...
  • Page 172: Diagnostic Information Via Light Emitting Diodes

    Diagnostics and troubleshooting Proline Promass E 300 EtherNet/IP Fault Possible causes Remedial action Parallel WLAN and Ethernet communication • Check network settings. • Temporarily enable only the WLAN as an interface. Web browser frozen and operation no longer Data transfer is active.
  • Page 173 Proline Promass E 300 EtherNet/IP Diagnostics and troubleshooting Color Meaning Flashing red A diagnostic event with "Warning" diagnostic behavior has occurred. A diagnostic event with "Alarm" diagnostic behavior has occurred. Flashing red/green The device restarts/self-test. Device status/module Flashes red slowly If >...
  • Page 174: Diagnostic Information On Local Display

    Diagnostics and troubleshooting Proline Promass E 300 EtherNet/IP 12.3 Diagnostic information on local display 12.3.1 Diagnostic message Faults detected by the self-monitoring system of the measuring device are displayed as a diagnostic message in alternation with the operational display. Operational display in alarm condition...
  • Page 175: Diagnostic Information

    Proline Promass E 300 EtherNet/IP Diagnostics and troubleshooting Diagnostic behavior Symbol Meaning Alarm • Measurement is interrupted. • Signal outputs and totalizers assume the defined alarm condition. • A diagnostic message is generated. Warning • Measurement is resumed. • The signal outputs and totalizers are not affected.
  • Page 176: Calling Up Remedial Measures

    Diagnostics and troubleshooting Proline Promass E 300 EtherNet/IP 12.3.2 Calling up remedial measures X X X X X X X X X X X X X X 20.50 S801 Supply voltage Menu Diagnostic list Diagnostics 1 S801 Supply voltage Diagnostics 2...
  • Page 177: Calling Up Remedy Information

    Proline Promass E 300 EtherNet/IP Diagnostics and troubleshooting A0031056 Status area with status signal Diagnostic information Remedial measures with service ID In addition, diagnostic events which have occurred can be shown in the Diagnostics menu: • Via parameter →  193 •...
  • Page 178 Diagnostics and troubleshooting Proline Promass E 300 EtherNet/IP Xxxxxx/…/…/ Device name: Xxxxxxx Mass flow: kg/h 12.34 Device tag: Xxxxxxx Volume flow: 12.34 m /h ³ Status signal: Function check (C) Xxxxxx Diagnostics 1: C485 Simu... Remedy information: Deactivate... Failure (F)
  • Page 179: Diagnostic Information Via Communication Interface

    Proline Promass E 300 EtherNet/IP Diagnostics and troubleshooting The user is in the Diagnostics menu. 1. Call up the desired parameter. 2. On the right in the working area, mouse over the parameter.  A tool tip with remedy information for the diagnostic event appears.
  • Page 180: Diagnostic Of Sensor

    Diagnostics and troubleshooting Proline Promass E 300 EtherNet/IP 12.8.1 Diagnostic of sensor Diagnostic information Remedy instructions Coding of diagnostic information (hex) Short text 022 Temperature sensor defective 1. Check or replace sensor • 0x10000BE electronic module (ISEM) • 0x10000BF 2. If available: Check connection •...
  • Page 181: Diagnostic Of Electronic

    Proline Promass E 300 EtherNet/IP Diagnostics and troubleshooting Diagnostic information Remedy instructions Coding of diagnostic information (hex) Short text 140 Sensor signal asymmetrical 1. Check or replace sensor 0x80000CC electronic module (ISEM) 2. If available: Check connection Status signal cable between sensor and...
  • Page 182 Diagnostics and troubleshooting Proline Promass E 300 EtherNet/IP Diagnostic information Remedy instructions Coding of diagnostic information (hex) Short text 262 Sensor electronic connection faulty 1. Check or replace connection 0x1000149 cable between sensor electronic module (ISEM) and main Status signal...
  • Page 183 Proline Promass E 300 EtherNet/IP Diagnostics and troubleshooting Diagnostic information Remedy instructions Coding of diagnostic information (hex) Short text 276 I/O module 1 to n faulty 1. Restart device • 0x100007B 2. Change I/O module • 0x1000081 Status signal Diagnostic behavior...
  • Page 184 Diagnostics and troubleshooting Proline Promass E 300 EtherNet/IP Diagnostic information Remedy instructions Coding of diagnostic information (hex) Short text 372 Sensor electronic (ISEM) faulty 1. Restart device • 0x10000A1 2. Check if failure recurs • 0x10000C7 3. Replace sensor electronic module •...
  • Page 185: Diagnostic Of Configuration

    Proline Promass E 300 EtherNet/IP Diagnostics and troubleshooting Diagnostic information Remedy instructions Coding of diagnostic information (hex) Short text 383 Memory content 1. Restart device 0x100016E 2. Delete T-DAT via ' R eset device' parameter Status signal 3. Replace T-DAT...
  • Page 186 Diagnostics and troubleshooting Proline Promass E 300 EtherNet/IP Diagnostic information Remedy instructions Coding of diagnostic information (hex) Short text 412 Processing download Download active, please wait 0x2000204 Status signal Diagnostic behavior Warning Diagnostic information Remedy instructions Coding of diagnostic information (hex)
  • Page 187 Proline Promass E 300 EtherNet/IP Diagnostics and troubleshooting Diagnostic information Remedy instructions Coding of diagnostic information (hex) Short text 443 Pulse output 1 to n 1. Check process • 0x800008C 2. Check pulse output settings • 0x8000121 Status signal Diagnostic behavior...
  • Page 188 Diagnostics and troubleshooting Proline Promass E 300 EtherNet/IP Diagnostic information Remedy instructions Coding of diagnostic information (hex) Short text 491 Current output 1 to n simulation Deactivate simulation 0x200000E Status signal Diagnostic behavior Warning Diagnostic information Remedy instructions Coding of diagnostic...
  • Page 189: Diagnostic Of Process

    Proline Promass E 300 EtherNet/IP Diagnostics and troubleshooting Diagnostic information Remedy instructions Coding of diagnostic information (hex) Short text 520 I/O 1 to n hardware configuration invalid 1. Check I/O hardware 0x1000276 configuration 2. Replace wrong I/O module Status signal 3.
  • Page 190 Diagnostics and troubleshooting Proline Promass E 300 EtherNet/IP Diagnostic information Remedy instructions Coding of diagnostic information (hex) Short text 830 Sensor temperature too high Reduce ambient temp. around the 0x80000C0 sensor housing Status signal Diagnostic behavior [from the Warning factory] Diagnostic behavior can be changed.
  • Page 191 Proline Promass E 300 EtherNet/IP Diagnostics and troubleshooting Diagnostic information Remedy instructions Coding of diagnostic information (hex) Short text 835 Process temperature too low Increase process temperature 0x80000C6 Status signal Diagnostic behavior [from the Warning factory] Diagnostic behavior can be changed.
  • Page 192 Diagnostics and troubleshooting Proline Promass E 300 EtherNet/IP Diagnostic information Remedy instructions Coding of diagnostic information (hex) Short text 913 Medium unsuitable 1. Check process conditions 0x80000CD 2. Check electronic modules or sensor Status signal Diagnostic behavior [from the Alarm factory] Diagnostic behavior can be changed.
  • Page 193: Pending Diagnostic Events

    Proline Promass E 300 EtherNet/IP Diagnostics and troubleshooting 12.9 Pending diagnostic events The Diagnostics menu allows the user to view the current diagnostic event and the previous diagnostic event separately. To call up the measures to rectify a diagnostic event: •...
  • Page 194: 12.11 Event Logbook

    Diagnostics and troubleshooting Proline Promass E 300 EtherNet/IP / ../Diagnose list Diagnostics F273 Main electronic Diagnostics 2 Diagnostics 3 A0014006-EN  32 Using the example of the local display To call up the measures to rectify a diagnostic event: • Via local display →  176 •...
  • Page 195: 12.11.2 Filtering The Event Logbook

    Proline Promass E 300 EtherNet/IP Diagnostics and troubleshooting 12.11.2 Filtering the event logbook Using the Filter options parameter you can define which category of event message is displayed in the Events list submenu. Navigation path Diagnostics → Event logbook → Filter options Filter categories •...
  • Page 196: 12.12 Resetting The Measuring Device

    Diagnostics and troubleshooting Proline Promass E 300 EtherNet/IP Info number Info name I1451 Monitoring on I1457 Measured error verification failed I1459 I/O module verification failed I1460 HBSI verification failed I1461 Sensor verification failed I1462 Sensor electronic module verific. failed I1512...
  • Page 197: 12.13 Device Information

    Proline Promass E 300 EtherNet/IP Diagnostics and troubleshooting Options Description Restart device The restart resets every parameter with data stored in volatile memory (RAM) to the factory setting (e.g. measured value data). The device configuration remains unchanged. Restore S-DAT backup Restores the data that is saved on the S-DAT.
  • Page 198: 12.14 Firmware History

    "Manufacturer' s information" document. The manufacturer' s information is available: • In the Download Area of the Endress+Hauser web site: www.endress.com → Downloads • Specify the following details: •...
  • Page 199: Maintenance

    • Observe the maximum permitted medium temperature for the measuring device . 13.2 Measuring and test equipment Endress+Hauser offers a variety of measuring and testing equipment, such as Netilion or device tests. Your Endress+Hauser Sales Center can provide detailed information on the services.
  • Page 200: Repair

    • The measuring devices have a modular design. • Spare parts are grouped into logical kits with the associated Installation Instructions. • Repairs are carried out by Endress+Hauser Service or by appropriately trained customers. • Certified devices can only be converted to other certified devices by Endress+Hauser Service or at the factory.
  • Page 201: Disposal

    Proline Promass E 300 EtherNet/IP Repair 14.5 Disposal If required by the Directive 2012/19/EU on waste electrical and electronic equipment (WEEE), the product is marked with the depicted symbol in order to minimize the disposal of WEEE as unsorted municipal waste. Do not dispose of products bearing this marking as unsorted municipal waste.
  • Page 202: Accessories

    Various accessories, which can be ordered with the device or subsequently from Endress +Hauser, are available for the device. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 203: For The Sensor

     If using oil as a heating medium, please consult with Endress+Hauser. Heating jackets cannot be used with sensors fitted with a rupture disk. Use the order code with the product root DK8003.
  • Page 204: Service-Specific Accessories

    Via the Internet: https://portal.endress.com/webapp/applicator Netilion lloT ecosystem: Unlock knowledge With the Netilion IIoT ecosystem,Endress+Hauser allows you to optimize your plant performance, digitize workflows, share knowledge, and enhance collaboration. Drawing upon decades of experience in process automation, Endress+Hauser offers the process industry an IIoT ecosystem designed to effortlessly extract insights from data.
  • Page 205: Technical Data

    Proline Promass E 300 EtherNet/IP Technical data Technical data 16.1 Application The measuring device is intended only for the flow measurement of liquids and gases. Depending on the version ordered, the measuring device can also measure potentially explosive, flammable, poisonous and oxidizing media.
  • Page 206: Input

    Technical data Proline Promass E 300 EtherNet/IP 16.3 Input Measured variable Direct measured variables • Mass flow • Density • Temperature Calculated measured variables • Volume flow • Corrected volume flow • Reference density Measuring range Measuring range for liquids Measuring range full scale values ...
  • Page 207 • Operating pressure to increase measurement accuracy (Endress+Hauser recommends the use of a pressure measuring device for absolute pressure, e.g. Cerabar M or Cerabar •...
  • Page 208 Technical data Proline Promass E 300 EtherNet/IP Open-circuit voltage ≤ 28.8 V (active) Possible input variables • Pressure • Temperature • Density Status input Maximum input values • DC –3 to 30 V • If status input is active (ON): R >3 kΩ...
  • Page 209: Output

    Proline Promass E 300 EtherNet/IP Technical data 16.4 Output Output signal EtherNet/IP Standards In accordance with IEEE 802.3 Current output 4 to 20 mA Signal mode Can be set to: • Active • Passive Current range Can be set to: •...
  • Page 210 Technical data Proline Promass E 300 EtherNet/IP Damping Configurable: 0 to 999 s Assignable measured • Mass flow variables • Volume flow • Corrected volume flow • Density • Reference density • Temperature • Electronics temperature • Oscillation frequency 0 •...
  • Page 211 Proline Promass E 300 EtherNet/IP Technical data Assignable measured • Mass flow variables • Volume flow • Corrected volume flow • Density • Reference density • Temperature • Electronics temperature • Oscillation frequency 0 • Oscillation damping 0 • Signal asymmetry •...
  • Page 212 Technical data Proline Promass E 300 EtherNet/IP Maximum switching • DC 30 V, 0.1 A capacity (passive) • AC 30 V, 0.5 A Assignable functions • Disable • On • Diagnostic behavior • Limit • Mass flow • Volume flow •...
  • Page 213 Proline Promass E 300 EtherNet/IP Technical data Pulse/frequency/switch output Pulse output Fault mode Choose from: • Actual value • No pulses Frequency output Fault mode Choose from: • Actual value • 0 Hz • Definable value between: 2 to 12 500 Hz...
  • Page 214: Light Emitting Diodes (Led)

    Technical data Proline Promass E 300 EtherNet/IP Light emitting diodes (LED) Status information Status indicated by various light emitting diodes The following information is displayed depending on the device version: • Supply voltage active • Data transmission active • Device alarm/error has occurred •...
  • Page 215: Power Supply

    Proline Promass E 300 EtherNet/IP Technical data 16.5 Power supply Terminal assignment →  34 Available device plugs →  34 Available device plugs →  34 Supply voltage Order code Terminal voltage Frequency range "Power supply" Option D DC 24 V ±20%...
  • Page 216: Performance Characteristics

    Technical data Proline Promass E 300 EtherNet/IP Cable entries • Cable gland: M20 × 1.5 with cable ⌀ 6 to 12 mm (0.24 to 0.47 in) • Thread for cable entry: • NPT ½" • G ½" • M20 • Device plug for digital communication: M12 Cable specification →...
  • Page 217: Zero Point Stability

    Proline Promass E 300 EtherNet/IP Technical data Zero point stability Zero point stability [mm] [in] [kg/h] [lb/min] ³⁄₈ 0.20 0.007 ½ 0.65 0.024 1.80 0.066 1½ 4.50 0.165 0.257 18.0 0.6615 Flow values Flow values as turndown parameters depending on nominal diameter.
  • Page 218 Technical data Proline Promass E 300 EtherNet/IP Repeatability o.r. = of reading; 1 g/cm = 1 kg/l; T = medium temperature Base repeatability Design fundamentals →  219 Mass flow and volume flow (liquids) ±0.075 % o.r. ±0.05 % o.r. (calibration option, for mass flow) Mass flow (gases) ±0.25 % o.r.
  • Page 219 Proline Promass E 300 EtherNet/IP Technical data [kg/m ] [°C] 40 80 120 160 200 240 280 320 [°F] -40 0 A0016609  34 Field density adjustment, for example at +20 °C (+68 °F) Temperature ±0.005 · T °C (± 0.005 · (T – 32) °F)
  • Page 220: Mounting

    The device is suitable for use outdoors and indoors with a relative humidity of 4 to 95 %. Operating height According to EN 61010-1 • ≤ 2 000 m (6 562 ft) • > 2 000 m (6 562 ft) with additional overvoltage protection (e.g. Endress+Hauser HAW Series) Endress+Hauser...
  • Page 221: Process

    Proline Promass E 300 EtherNet/IP Technical data Degree of protection Transmitter • IP66/67, Type 4X enclosure, suitable for pollution degree 4 • When the housing is open: IP20, Type 1 enclosure, suitable for pollution degree 2 • Display module: IP20, Type 1 enclosure, suitable for pollution degree 2 Optional Order code for "Sensor options", option CM "IP69"...
  • Page 222: Flow Limit

    Technical data Proline Promass E 300 EtherNet/IP Pressure-temperature For an overview of the pressure-temperature ratings for the process connections, see ratings the Technical Information Sensor housing The sensor housing is filled with dry nitrogen gas and protects the electronics and mechanics inside.
  • Page 223: 16.10 Mechanical Construction

    Proline Promass E 300 EtherNet/IP Technical data • The minimum recommended full scale value is approx. 1/20 of the maximum full scale value • In most applications, 20 to 50 % of the maximum full scale value can be considered ideal •...
  • Page 224: Transmitter Housing

    Technical data Proline Promass E 300 EtherNet/IP Weight [lbs] [in] Materials Transmitter housing Order code for "Housing": • Option A "Aluminum, coated": aluminum, AlSi10Mg, coated • Option B "Stainless, hygienic": stainless steel, 1.4404 (316L) Window material Order code for "Housing": •...
  • Page 225: Process Connections

    Proline Promass E 300 EtherNet/IP Technical data Process connections • Flanges according to EN 1092-1 (DIN2501) / according to ASME B 16.5 / as per JIS B2220: Stainless steel, 1.4404 (F316/F316L) • All other process connections: Stainless steel, 1.4404 (316/316L) Available process connections→...
  • Page 226: 16.11 Operability

    Technical data Proline Promass E 300 EtherNet/IP Category Method Option(s) order code "Measuring tube mat., wetted surface" Ra ≤ 0.38 µm (15 µin) Mechanically polished Ra ≤ 0.38 µm (15 µin) Mechanically polished , welds in as-welded condition Ra according to ISO 21920 Except for inaccessible welds between pipe and manifold 16.11 Operability...
  • Page 227 Proline Promass E 300 EtherNet/IP Technical data Via remote display and operating module DKX001 The remote display and operating module DKX001 is available as an optional extra →  202.. • The remote display and operating module DKX001 is only available for the following housing version: order code for "Housing": option A "Aluminum, coated"...
  • Page 228 Technical data Proline Promass E 300 EtherNet/IP Supported operating tools Different operating tools can be used for local or remote access to the measuring device. Depending on the operating tool used, access is possible with different operating units and via a variety of interfaces.
  • Page 229: Historom Data Management

    Proline Promass E 300 EtherNet/IP Technical data • Export parameter settings (.csv file or PDF file, document the measuring point configuration) • Export the Heartbeat Technology verification report (PDF file, only available with the Heartbeat Verification →  232 application package) •...
  • Page 230: 16.12 Certificates And Approvals

    The device meets the legal requirements of the applicable EU Directives. These are listed in the corresponding EU Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. UKCA marking The device meets the legal requirements of the applicable UK regulations (Statutory Instruments).
  • Page 231 Directive a) PED/G1/x (x = category) or b) PESR/G1/x (x = category) on the sensor nameplate, Endress+Hauser confirms compliance with the "Essential Safety Requirements" a) specified in Annex I of the Pressure Equipment Directive 2014/68/EU or b) Schedule 2 of Statutory Instruments 2016 No. 1105.
  • Page 232: 16.13 Application Packages

    The application packages can be ordered with the device or subsequently from Endress+Hauser. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 233: Heartbeat Verification

    Proline Promass E 300 EtherNet/IP Technical data Event log: Memory volume is extended from 20 message entries (standard version) to up to 100 entries. Data logging (line recorder): • Memory capacity for up to 1000 measured values is activated. • 250 measured values can be output via each of the 4 memory channels. The recording interval can be defined and configured by the user.
  • Page 234: 16.14 Accessories

    For an overview of the scope of the associated Technical Documentation, refer to the following: • Device Viewer (www.endress.com/deviceviewer): Enter the serial number from the nameplate • Endress+Hauser Operations app: Enter serial number from nameplate or scan matrix code on nameplate. Standard documentation Brief Operating Instructions...
  • Page 235 Proline Promass E 300 EtherNet/IP Technical data Description of Device Parameters Documentation code PROFINET with FOUNDATIO PROFIBUS PROFIBUS Modbus Ethernet- Measuring device HART N Fieldbus RS485 EtherNet/IP PROFINET Promass 300 GP01057D GP01094D GP01058D GP01134D GP01059D GP01114D GP01115D GP01168D Supplementary device-...
  • Page 236: Installation Instructions

    Technical data Proline Promass E 300 EtherNet/IP Contents Documentation code Concentration measurement SD02004 Petroleum SD02096D Installation instructions Contents Note Installation instructions for spare part sets and • Access the overview of all the available spare part accessories sets via Device Viewer →  200 •...
  • Page 237: Index

    Proline Promass E 300 EtherNet/IP Index Index 0 … 9 Closing ....... . . 57 Explanation .
  • Page 238 Index Proline Promass E 300 EtherNet/IP Display values Flow direction ......22, 27 For locking status ......149 Flow limit .
  • Page 239 Proline Promass E 300 EtherNet/IP Index Measurement accuracy ..... . 216 Measuring and test equipment ....199...
  • Page 240 Index Proline Promass E 300 EtherNet/IP Reset access code (Submenu) ....141 Triggering pressure ..... . 222 Sensor adjustment (Submenu) .
  • Page 241 Proline Promass E 300 EtherNet/IP Index Structure Measuring device ......14 Technical data, overview ..... 205...
  • Page 242 Index Proline Promass E 300 EtherNet/IP Pulse/frequency/switch output ..105, 107, 111 Relay output 1 to n ..... . . 113 Status input 1 to n .
  • Page 244 *71675064* 71675064 www.addresses.endress.com...
Save PDF