Working Technique
The nozzle length enables cleaning pockets over
5 mm deep.
Never treat with AIR-FLOW handy 3.0
PERIO the pockets down to the apex. At least
3 mm of bone must be present to prevent the
risk of emphysema. Verify with radiographs
pre-treatment.
Do not use the PERIO-FLOW method
immediately after a sub gingival scaling or a root
planning treatment.
34
FB-543_4_rev_B_Handy 3 Perio.indd 34
FB-543_4_rev_B_Handy 3 Perio.indd 34
Radna tehnika
Duljina mlaznice omogućuje čišćenje džepova
dubljih od 5 mm.
Nikad ne tretirajte džepove sve do vrha
korijena jedinicom AIR-FLOW handy 3.0 PERIO.
Mora postojati najmanje 3 mm kosti kako bi se
spriječio rizik od emfizema. Potvrdite prethodnim
radiografskim snimanjem.
Ne koristite metodu PERIO-FLOW
neposredno nakon uklanjanja subgingivalnog
kamenca ili kiretaže.
Behandeltechniek
De lengte van het spuitstuk maakt reiniging van
pockets dieper dan 5 mm mogelijk.
Behandel de pockets met de AIR-FLOW
handy 3.0 PERIO nooit helemaal tot aan de apex.
Er moet tenminste 3 mm bot aanwezig zijn om
risico op emfyseem te vermijden. Controleer de
voorbehandeling met röntgenfoto's.
Gebruik de PERIO-FLOW methode niet
direct na een sub-gingivale scaling of root planing
behandeling.